Journal of applied clinical medical physics 20080128 1
An essential component of quality assurance in radiation therapy is verifying the accuracy of monitor unit calculations. For tangential breast fields, monitor unit differences between primary calculations and second checks are usually larger than considered acceptable at other anatomical sites. A simple model to reconcile the differences between sophisticated and simple algorithms is presented, based on estimating the the volume irradiated by the field, replacing the breast contour with a rectan ...[more]