Unknown

Dataset Information

0

No Difference in Conventional Polyethylene Wear Between Yttria-stabilized Zirconia and Cobalt-chromium-molybdenum Femoral Heads at 10 Years.


ABSTRACT: Background:Concerns have arisen regarding deterioration of wear properties of yttria-stabilized zirconia (YSZ) femoral head on conventional polyethylene (PE) bearings due to YSZ phase transformation. Questions/Purposes:The purpose of this study was to determine if there is a difference in long-term PE wear properties between YSZ and cobalt-chromium-molybdenum (Co-Cr-Mo) femoral heads. Methods:Ten-year radiographic wear assessment was performed on a cohort of patients enrolled in a prospective randomized clinical trial comparing total hip arthroplasty with YSZ or Co-Cr-Mo femoral heads on conventional, non-cross-linked PE. Results:PE linear wear, annualized wear, and steady-state wear rates remained low and similar between groups. No cases of osteolysis were observed. Conclusions:Measured conventional PE wear was similar between YSZ and Co-Cr-Mo femoral heads with the steady-state wear rates for both remaining below the generally accepted threshold at which osteolysis typically occurs. Whether clinically relevant phase transformation with YSZ femoral heads occurs is uncertain; however, the use of YSZ femoral heads in this study was not associated with increased PE wear, osteolysis, or deterioration of wear properties.

SUBMITTER: Morrison TA 

PROVIDER: S-EPMC5786591 | biostudies-literature | 2018 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

No Difference in Conventional Polyethylene Wear Between Yttria-stabilized Zirconia and Cobalt-chromium-molybdenum Femoral Heads at 10 Years.

Morrison Todd A TA   Moore Rebecca D RD   Meng Jia J   Rimnac Clare M CM   Kraay Matthew J MJ  

HSS journal : the musculoskeletal journal of Hospital for Special Surgery 20171013 1


<h4>Background</h4>Concerns have arisen regarding deterioration of wear properties of yttria-stabilized zirconia (YSZ) femoral head on conventional polyethylene (PE) bearings due to YSZ phase transformation.<h4>Questions/purposes</h4>The purpose of this study was to determine if there is a difference in long-term PE wear properties between YSZ and cobalt-chromium-molybdenum (Co-Cr-Mo) femoral heads.<h4>Methods</h4>Ten-year radiographic wear assessment was performed on a cohort of patients enroll  ...[more]

Similar Datasets

| S-EPMC2553166 | biostudies-other
| S-EPMC6898179 | biostudies-literature
| S-EPMC7339873 | biostudies-literature
| S-EPMC4291567 | biostudies-other
| S-EPMC4148670 | biostudies-other