Scientific reports 20180205 1
Variability in the x-ray tube current used in computed tomography may affect quantitative features extracted from the images. To investigate these effects, we scanned the Credence Cartridge Radiomics phantom 12 times, varying the tube current from 25 to 300 mA∙s while keeping the other acquisition parameters constant. For each of the scans, we extracted 48 radiomic features from the categories of intensity histogram (n = 10), gray-level run length matrix (n = 11), gray-level co-occurrence matrix ...[more]