Unknown

Dataset Information

0

Performance of the BioPlex 2200 HIV Ag-Ab assay for identifying acute HIV infection.


ABSTRACT: BACKGROUND:Assays that detect HIV antigen (Ag) and antibody (Ab) can be used to screen for HIV infection. OBJECTIVES:To compare the performance of the BioPlex 2200 HIV Ag-Ab assay and two other Ag/Ab combination assays for detection of acute HIV infection. STUDY DESIGN:Samples were obtained from 24 individuals (18 from the US, 6 from South Africa); these individuals were classified as having acute infection based on the following criteria: positive qualitative RNA assay; two negative rapid tests; negative discriminatory test. The samples were tested with the BioPlex assay, the ARCHITECT HIV Ag/Ab Combo test, the Bio-Rad GS HIV Combo Ag-Ab EIA test, and a viral load assay. RESULTS:Twelve (50.0%) of 24 samples had RNA detected only (?>?40 to 13,476 copies/mL). Ten (43.5%) samples had reactive results with all three Ag/Ab assays, one sample was reactive with the ARCHITECT and Bio-Rad assays, and one sample was reactive with the Bio-Rad and BioPlex assays. The 11 samples that were reactive with the BioPlex assay had viral loads from 83,010 to >750,000 copies/mL; 9/11 samples were classified as Ag positive/Ab negative by the BioPlex assay. CONCLUSIONS:Detection of acute HIV infection was similar for the BioPlex assay and two other Ag/Ab assays. All three tests were less sensitive than a qualitative RNA assay and only detected HIV Ag when the viral load was high. The BioPlex assay detected acute infection in about half of the cases, and identified most of those infections as Ag positive/Ab negative.

SUBMITTER: Eshleman SH 

PROVIDER: S-EPMC5807223 | biostudies-literature | 2018 Feb - Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Performance of the BioPlex 2200 HIV Ag-Ab assay for identifying acute HIV infection.

Eshleman Susan H SH   Piwowar-Manning Estelle E   Sivay Mariya V MV   Debevec Barbara B   Veater Stephanie S   McKinstry Laura L   Bekker Linda-Gail LG   Mannheimer Sharon S   Grant Robert M RM   Chesney Margaret A MA   Coates Thomas J TJ   Koblin Beryl A BA   Fogel Jessica M JM  

Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology 20180111


<h4>Background</h4>Assays that detect HIV antigen (Ag) and antibody (Ab) can be used to screen for HIV infection.<h4>Objectives</h4>To compare the performance of the BioPlex 2200 HIV Ag-Ab assay and two other Ag/Ab combination assays for detection of acute HIV infection.<h4>Study design</h4>Samples were obtained from 24 individuals (18 from the US, 6 from South Africa); these individuals were classified as having acute infection based on the following criteria: positive qualitative RNA assay; tw  ...[more]

Similar Datasets

| S-EPMC5863103 | biostudies-literature
| S-EPMC8682874 | biostudies-literature
| S-EPMC6926103 | biostudies-literature