Unknown

Dataset Information

0

Evaluation of Hematocrit Influence on Self-Monitoring of Blood Glucose Based on ISO 15197:2013: Comparison of a Novel System With Five Systems With Different Hematocrit Ranges.


ABSTRACT:

Introduction

ISO 15197:2013 recommends testing procedures and acceptance criteria for the evaluation of influence quantities such as hematocrit on measurement results with systems for self-monitoring of blood glucose (SMBG). In this study, hematocrit influence was evaluated for a novel SMBG system (system A) and five other systems with different hematocrit ranges based on ISO 15197:2013.

Methods

Test procedures were performed with one test strip lot for each system. Each system was tested within the hematocrit range indicated in the manufacturer's labeling (system A: 10-65%, B: 15-65%, C: 20-60%, D: 35-60%, E: 30-60%, F: 30-55%). According to ISO 15197:2013, clause 6.4.2, venous blood samples were used for the evaluation of hematocrit influence. The evaluation was performed for three glucose concentration categories (30-50 mg/dL, 96-144 mg/dL, and 280-420 mg/dL). For each glucose concentration category, at least five different hematocrit levels were investigated.

Results

The novel system A and systems B, E, and F complied with the tested lot with the defined criteria and showed ?10 mg/dL and ?10% difference between the test sample and the respective control sample with a hematocrit value of 42% ± 2% for BG concentrations <100 mg/dL and ?100 mg/dL, respectively. Two systems showed >10% difference at glucose concentrations ?100 mg/dL.

Conclusions

Remarkable hematocrit influence within the labeled hematocrit range was obtained in two systems with the tested reagent system lot. Adequate SMBG systems should be carefully chosen by patients and their health care professionals, particularly for patients with increased and decreased hematocrit values.

SUBMITTER: Hattemer A 

PROVIDER: S-EPMC5851235 | biostudies-literature | 2018 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of Hematocrit Influence on Self-Monitoring of Blood Glucose Based on ISO 15197:2013: Comparison of a Novel System With Five Systems With Different Hematocrit Ranges.

Hattemer Andrew A   Wardat Sami S  

Journal of diabetes science and technology 20180129 2


<h4>Introduction</h4>ISO 15197:2013 recommends testing procedures and acceptance criteria for the evaluation of influence quantities such as hematocrit on measurement results with systems for self-monitoring of blood glucose (SMBG). In this study, hematocrit influence was evaluated for a novel SMBG system (system A) and five other systems with different hematocrit ranges based on ISO 15197:2013.<h4>Methods</h4>Test procedures were performed with one test strip lot for each system. Each system wa  ...[more]

Similar Datasets

| S-EPMC6104257 | biostudies-literature
| S-EPMC7039612 | biostudies-literature
| S-EPMC4667333 | biostudies-literature
| S-EPMC5500346 | biostudies-literature
| S-EPMC6080122 | biostudies-literature
| S-EPMC5851223 | biostudies-literature
| S-EPMC6080115 | biostudies-literature
| S-EPMC6203032 | biostudies-literature
| S-EPMC7463700 | biostudies-literature
| S-EPMC3558677 | biostudies-literature