System accuracy evaluation of 18 CE-marked current-generation blood glucose monitoring systems based on EN ISO 15197:2015.


ABSTRACT: OBJECTIVE:Accuracy of 18 current-generation blood glucose monitoring systems (BGMS) available in Europe was evaluated applying criteria adapted from EN ISO 15197:2015 with one reagent system lot. BGMS were selected based on market research data. RESEARCH DESIGN AND METHODS:The BGMS ABRA, Accu-Chek Guide, AURUM, CareSens Dual, CERA-CHEK 1CODE, ContourNext One, eBsensor, FreeStyle Freedom Lite, GL50 evo, GlucoCheck GOLD, GlucoMen areo 2K, GluNEO, MyStar DoseCoach, OneTouch Verio Flex, Pic GlucoTest, Rightest GM700S, TRUEyou, and WaveSense JAZZ Wireless were tested using capillary blood from 100 different subjects and assessing the percentage of results within ±15 mg/dL (0.83 mmol/L) or 15% of comparison method results for BG concentrations below or above 100 mg/dL (5.55 mmol/L), respectively. In addition, the minimal deviation from comparison method results within which ?95% of results of the respective BGMS were found was calculated. RESULTS:In total, 14 BGMS had ?95% of results within ±15 mg/dL (0.83 mmol/L) or ±15% and 3 BGMS had ?95% of results within ±10 mg/dL (0.55 mmol/L) or ±10% of the results obtained with the comparison method. The smallest deviation from comparison method results within which ?95% of results were found was ±7.7 mg/dL (0.43 mmol/L) or ±7.7%; the highest deviation was ±19.7 mg/dL (1.09 mmol/L) or ±19.7%. CONCLUSIONS:This accuracy evaluation shows that not all CE-labeled BGMS fulfill accuracy requirements of ISO 15197 reliably and that there is considerable variation even among BGMS fulfilling these criteria. This safety-related information should be taken into account by patients and healthcare professionals when making therapy decisions. TRIAL REGISTRATION NUMBER:NCT03737188.

SUBMITTER: Pleus S 

PROVIDER: S-EPMC7039612 | biostudies-literature | 2020 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

System accuracy evaluation of 18 CE-marked current-generation blood glucose monitoring systems based on EN ISO 15197:2015.

Pleus Stefan S   Baumstark Annette A   Jendrike Nina N   Mende Jochen J   Link Manuela M   Zschornack Eva E   Haug Cornelia C   Freckmann Guido G  

BMJ open diabetes research & care 20200101 1


<h4>Objective</h4>Accuracy of 18 current-generation blood glucose monitoring systems (BGMS) available in Europe was evaluated applying criteria adapted from EN ISO 15197:2015 with one reagent system lot. BGMS were selected based on market research data.<h4>Research design and methods</h4>The BGMS ABRA, Accu-Chek Guide, AURUM, CareSens Dual, CERA-CHEK 1CODE, ContourNext One, eBsensor, FreeStyle Freedom Lite, GL50 evo, GlucoCheck GOLD, GlucoMen areo 2K, GluNEO, MyStar DoseCoach, OneTouch Verio Fle  ...[more]

Similar Datasets

| S-EPMC6104257 | biostudies-literature
| S-EPMC4667333 | biostudies-literature
| S-EPMC5851235 | biostudies-literature
| S-EPMC5500346 | biostudies-literature
| S-EPMC2959540 | biostudies-literature
| S-EPMC5561792 | biostudies-other
| S-EPMC5563027 | biostudies-literature
| S-EPMC2986222 | biostudies-literature
| S-EPMC8569926 | biostudies-literature
| S-EPMC3414094 | biostudies-literature