Validation of the doubly labeled water method using off-axis integrated cavity output spectroscopy and isotope ratio mass spectrometry.


ABSTRACT: When the doubly labeled water (DLW) method is used to measure total daily energy expenditure (TDEE), isotope measurements are typically performed using isotope ratio mass spectrometry (IRMS). New technologies, such as off-axis integrated cavity output spectroscopy (OA-ICOS) provide comparable isotopic measurements of standard waters and human urine samples, but the accuracy of carbon dioxide production (V?co2) determined with OA-ICOS has not been demonstrated. We compared simultaneous measurement V?co2 obtained using whole-room indirect calorimetry (IC) with DLW-based measurements from IRMS and OA-ICOS. Seventeen subjects (10 female; 22 to 63 yr) were studied for 7 consecutive days in the IC. Subjects consumed a dose of 0.25 g H218O (98% APE) and 0.14 g 2H2O (99.8% APE) per kilogram of total body water, and urine samples were obtained on days 1 and 8 to measure average daily V?co2 using OA-ICOS and IRMS. V?co2 was calculated using both the plateau and intercept methods. There were no differences in V?co2 measured by OA-ICOS or IRMS compared with IC when the plateau method was used. When the intercept method was used, V?co2 using OA-ICOS did not differ from IC, but V?co2 measured using IRMS was significantly lower than IC. Accuracy (~1-5%), precision (~8%), intraclass correlation coefficients ( R?=?0.87-90), and root mean squared error (30-40 liters/day) of V?co2 measured by OA-ICOS and IRMS were similar. Both OA-ICOS and IRMS produced measurements of V?co2 with comparable accuracy and precision compared with IC.

SUBMITTER: Melanson EL 

PROVIDER: S-EPMC5866416 | biostudies-literature | 2018 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Validation of the doubly labeled water method using off-axis integrated cavity output spectroscopy and isotope ratio mass spectrometry.

Melanson Edward L EL   Swibas Tracy T   Kohrt Wendy M WM   Catenacci Vicki A VA   Creasy Seth A SA   Plasqui Guy G   Wouters Loek L   Speakman John R JR   Berman Elena S F ESF  

American journal of physiology. Endocrinology and metabolism 20171003 2


When the doubly labeled water (DLW) method is used to measure total daily energy expenditure (TDEE), isotope measurements are typically performed using isotope ratio mass spectrometry (IRMS). New technologies, such as off-axis integrated cavity output spectroscopy (OA-ICOS) provide comparable isotopic measurements of standard waters and human urine samples, but the accuracy of carbon dioxide production (V̇co<sub>2</sub>) determined with OA-ICOS has not been demonstrated. We compared simultaneous  ...[more]

Similar Datasets

| S-EPMC8754563 | biostudies-literature
| S-EPMC4891929 | biostudies-literature
| S-EPMC4974350 | biostudies-literature
| S-EPMC7066053 | biostudies-literature
| S-EPMC6501184 | biostudies-literature
| S-EPMC4939407 | biostudies-literature
| S-EPMC3896219 | biostudies-literature
| S-EPMC6271514 | biostudies-literature
| S-EPMC9795337 | biostudies-literature
| S-EPMC4673621 | biostudies-literature