Automated flow injection method for the high precision determination of drift tube ion mobility collision cross sections.


ABSTRACT: The field of ion mobility-based omics studies requires high-quality collision cross section (CCS) libraries to effectively utilize CCS as a molecular descriptor. Absolute CCS values with the highest precision are obtained on drift tube instruments by measuring the drift time of ions at multiple drift voltages, commonly referred to as a 'stepped field' experiment. However, generating large scale absolute CCS libraries from drift tube instruments is time consuming due to the current lack of high-throughput methods. This communication reports a fully automated stepped-field method to acquire absolute CCS on commercially available equipment. Using a drift tube ion mobility-mass spectrometer (DTIM-MS) coupled to a minimally modified liquid chromatography (LC) system, CCS values can be measured online with a carefully timed flow injection analysis (FIA) experiment. Results demonstrate that the FIA stepped-field method yields CCS values which are of high analytical precision (<0.4% relative standard deviation, RSD) and accuracy (?0.4% difference) comparable to CCS values obtained using traditional direct-infusion stepped-field experiments. This high-throughput CCS method consumes very little sample volume (20 ?L) and will expedite the generation of large-scale CCS libraries to support molecular identification within global untargeted studies.

SUBMITTER: Nichols CM 

PROVIDER: S-EPMC5869168 | biostudies-literature | 2018 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Automated flow injection method for the high precision determination of drift tube ion mobility collision cross sections.

Nichols Charles M CM   May Jody C JC   Sherrod Stacy D SD   McLean John A JA  

The Analyst 20180301 7


The field of ion mobility-based omics studies requires high-quality collision cross section (CCS) libraries to effectively utilize CCS as a molecular descriptor. Absolute CCS values with the highest precision are obtained on drift tube instruments by measuring the drift time of ions at multiple drift voltages, commonly referred to as a 'stepped field' experiment. However, generating large scale absolute CCS libraries from drift tube instruments is time consuming due to the current lack of high-t  ...[more]

Similar Datasets

| S-EPMC5744684 | biostudies-literature
| S-EPMC3931330 | biostudies-literature
| S-EPMC7289629 | biostudies-literature
| S-EPMC5744691 | biostudies-literature
| S-EPMC4004193 | biostudies-other
| S-EPMC5853271 | biostudies-literature
| S-EPMC6628689 | biostudies-literature
| S-EPMC6520154 | biostudies-literature
| S-EPMC10516336 | biostudies-literature
| S-EPMC10103148 | biostudies-literature