Translating bed total body irradiation lung shielding and dose optimization using asymmetric MLC apertures.


ABSTRACT: A revised translating bed total body irradiation (TBI) technique is developed for shielding organs at risk (lungs) to tolerance dose limits, and optimizing dose distribution in three dimensions (3D) using an asymmetrically-adjusted, dynamic multileaf collimator. We present a dosimetric comparison of this technique with a previously developed symmetric MLC-based TBI technique. An anthropomor-phic RANDO phantom is CT scanned with 3 mm slice thickness. Radiological depths (RD) are calculated on individual CT slices along the divergent ray lines. Asymmetric MLC apertures are defined every 9 mm over the phantom length in the craniocaudal direction. Individual asymmetric MLC leaf positions are optimized based on RD values of all slices for uniform dose distributions. Dose calculations are performed in the Eclipse treatment planning system over these optimized MLC apertures. Dose uniformity along midline of the RANDO phantom is within the confidence limit (CL) of 2.1% (with a confidence probability p = 0.065). The issue of over- and underdose at the interfaces that is observed when symmetric MLC apertures are used is reduced from more than ± 4% to less than ± 1.5% with asymmetric MLC apertures. Lungs are shielded by 20%, 30%, and 40% of the prescribed dose by adjusting the MLC apertures. Dose-volume histogram analysis confirms that the revised technique provides effective lung shielding, as well as a homogeneous dose coverage to the whole body. The asymmetric technique also reduces hot and cold spots at lung-tissue interfaces compared to previous symmetric MLC-based TBI technique. MLC-based shielding of OARs eliminates the need to fabricate and setup cumbersome patient-specific physical blocks.

SUBMITTER: Ahmed S 

PROVIDER: S-EPMC5875554 | biostudies-literature | 2016 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Translating bed total body irradiation lung shielding and dose optimization using asymmetric MLC apertures.

Ahmed Shahbaz S   Brown Derek D   Ahmed Saad B S SB   Kakakhel Muhammad B MB   Muhammad Wazir W   Hussain Amjad A  

Journal of applied clinical medical physics 20160308 2


A revised translating bed total body irradiation (TBI) technique is developed for shielding organs at risk (lungs) to tolerance dose limits, and optimizing dose distribution in three dimensions (3D) using an asymmetrically-adjusted, dynamic multileaf collimator. We present a dosimetric comparison of this technique with a previously developed symmetric MLC-based TBI technique. An anthropomor-phic RANDO phantom is CT scanned with 3 mm slice thickness. Radiological depths (RD) are calculated on ind  ...[more]

Similar Datasets

2010-06-09 | E-GEOD-1366 | biostudies-arrayexpress
2004-06-17 | GSE1366 | GEO
| S-EPMC8315067 | biostudies-literature
| S-EPMC5718217 | biostudies-literature
| S-EPMC4844345 | biostudies-literature
| S-EPMC7807637 | biostudies-literature
| S-EPMC6554899 | biostudies-other
| S-EPMC6598379 | biostudies-literature
| S-EPMC6380561 | biostudies-literature
| S-EPMC6401363 | biostudies-literature