Unknown

Dataset Information

0

Real-world experience with a Paclitaxel-Coated Balloon for the treatment of atherosclerotic infrainguinal arteries: 12-month interim results of the BIOLUX P-III registry first year of enrolment.


ABSTRACT: Background:Endovascular management of atherosclerotic infrainguinal arteries recently shifted towards drug eluting devices, designed to locally prevent the restenosis process. Numerous clinical studies report an advantage of drug coated balloons over uncoated balloon angioplasty in treating lower extremity peripheral artery disease. However, as coating and balloon platforms are different, each device requires dedicated clinical evaluations. Objective:The aim of the study is to further investigate the safety and effectiveness of a Paclitaxel-Coated Balloon for the treatment of atherosclerotic infrainguinal arteries in a real-world setting. Methods:203 patients out of a final sample of 882 were enrolled in this prospective multicenter, observational, all-comers registry during the first 12 months. The primary endpoints were major adverse events (defined as procedure or device related death within 30 days post index procedure, clinically-driven target lesion revascularization or major target limb amputation) at 6 months and freedom from clinically-driven target lesion revascularization at 12 months. Both endpoints were adjudicated by a Clinical Events Committee. Results:Mean patient age was 70.2±10.4 years (60.1% male). 47.3% of the patients were diabetic and 67.5% had a history of smoking. Severe claudication was reported in 37.4% and 40% had critical limb ischemia. 257 lesions, including 13.2% in the infrapopliteal territory, were treated with Passeo-18 Lux (mean lesion length 75.1 mm±69.4, 20% occlusions, 76.3% calcified). At 6 months, the rate of major adverse events was 5.5% (95%CI 3.1-9.7). Freedom from clinically-driven target lesion revascularization at 12 months was 93.2% (95%CI 89.1-95.8). All causes mortality was 6.5% (95%CI 3.8-11.0) and overall amputation rate was 4.2% (95%CI 2.1-8.3) at 12 months. Conclusion:In a real-world environment, the BIOLUX P-III registry preliminary results confirm the safety and efficacy of the Paclitaxel-Coated Passeo-18 Lux balloon as a stand-alone treatment option for atherosclerotic infrainguinal arteries.

SUBMITTER: Brodmann M 

PROVIDER: S-EPMC5944304 | biostudies-literature | 2017 Oct-Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Real-world experience with a Paclitaxel-Coated Balloon for the treatment of atherosclerotic infrainguinal arteries: 12-month interim results of the BIOLUX P-III registry first year of enrolment.

Brodmann Marianne M   Zeller Thomas T   Christensen Johnny J   Binkert Christoph C   Spak Lubomir L   Schröder Henrik H   Righini Paolo P   Nano Giovanni G   Tepe Gunnar G  

Jornal vascular brasileiro 20171001 4


<h4>Background</h4>Endovascular management of atherosclerotic infrainguinal arteries recently shifted towards drug eluting devices, designed to locally prevent the restenosis process. Numerous clinical studies report an advantage of drug coated balloons over uncoated balloon angioplasty in treating lower extremity peripheral artery disease. However, as coating and balloon platforms are different, each device requires dedicated clinical evaluations.<h4>Objective</h4>The aim of the study is to fur  ...[more]

Similar Datasets

| S-EPMC7806550 | biostudies-literature
| S-EPMC7082893 | biostudies-literature
| S-EPMC6010649 | biostudies-literature
| S-EPMC4733051 | biostudies-literature
| S-EPMC5951204 | biostudies-literature
| S-EPMC5610565 | biostudies-literature
| S-EPMC8460758 | biostudies-literature
| S-EPMC6455750 | biostudies-literature
| S-EPMC5084407 | biostudies-literature
| S-EPMC5459585 | biostudies-literature