Consistent tracer administration profile improves test-retest repeatability of myocardial blood flow quantification with 82Rb dynamic PET imaging.


ABSTRACT:

Objectives

Quantification of myocardial blood flow (MBF) and stress/rest flow reserve is used increasingly to diagnose multi-vessel coronary artery disease and micro-vascular disease with PET imaging. However, variability in the measurements may limit physician confidence to direct revascularization therapies based on specific threshold values. This study evaluated the effects of rubidium-82 (82Rb) tracer injection profile using a constant-activity-rate (CA) vs a constant-flow-rate (CF) infusion to improve test-retest repeatability of MBF measurements.

Method

22 participants underwent single-session 82Rb dynamic PET imaging during rest and dipyridamole stress using one of 2 test-retest infusion protocols: CA-CA (n = 12) or CA-CF (n = 10). MBF was quantified using a single-tissue-compartment model (1TCM) and a simplified retention model (SRM). Non-parametric test-retest repeatability coefficients (RPCnp) were compared between groups. Myocardium-to-blood contrast and signal-to-noise ratios of the late uptake images (2 to 6 minutes) were also compared to evaluate standard myocardial perfusion image (MPI) quality.

Results

MBF values in the CA-CA group were more repeatable (smaller RPCnp) than the CA-CF group using the 1TCM at rest alone, rest and stress combined, and stress/rest reserve (21% vs 36%, 16% vs 19%, and 20% vs 27%, P < 0.05, respectively), and using the SRM at Rest and Stress alone, Rest and Stress combined, and stress/rest reserve (21% vs 38%, 15% vs 25%, 22% vs 38%, and 23% vs 49%, P < 0.05, respectively). In terms of image quality, myocardium-to-blood contrast and signal-to-noise ratios were not significantly different between groups.

Conclusions

Constant-activity-rate 'square-wave' infusion of 82Rb produces more repeatable tracer injection profiles and decreases the test-retest variability of MBF measurements, when compared to a constant-flow-rate 'bolus' administration of 82Rb, especially with SRM, and without compromising standard MPI quality.

SUBMITTER: Klein R 

PROVIDER: S-EPMC5966478 | biostudies-literature | 2018 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Consistent tracer administration profile improves test-retest repeatability of myocardial blood flow quantification with <sup>82</sup>Rb dynamic PET imaging.

Klein Ran R   Ocneanu Adrian A   Renaud Jennifer M JM   Ziadi Maria C MC   Beanlands Rob S B RSB   deKemp Robert A RA  

Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology 20180601 3


<h4>Objectives</h4>Quantification of myocardial blood flow (MBF) and stress/rest flow reserve is used increasingly to diagnose multi-vessel coronary artery disease and micro-vascular disease with PET imaging. However, variability in the measurements may limit physician confidence to direct revascularization therapies based on specific threshold values. This study evaluated the effects of rubidium-82 (<sup>82</sup>Rb) tracer injection profile using a constant-activity-rate (CA) vs a constant-flow  ...[more]

Similar Datasets

| S-EPMC6826855 | biostudies-literature
| S-EPMC7034080 | biostudies-literature
| S-EPMC5331165 | biostudies-literature
| S-EPMC5807225 | biostudies-literature
| S-EPMC8786991 | biostudies-literature
| S-EPMC7705644 | biostudies-literature
| S-EPMC4260449 | biostudies-literature
| S-EPMC8138337 | biostudies-literature
| S-EPMC6504625 | biostudies-literature
| S-EPMC6151305 | biostudies-literature