ABSTRACT: Background
The diagnosis of dengue is complex. Until recently, only specialised laboratories were able to confirm dengue infection. However, this has changed with the newly available immunochromatographic rapid tests. Early diagnosis is of great interest, and point-of-care rapid tests have been increasingly used in Brazil. Most of those tests have not undergone validation in the Brazilian population. In this context, we decided to evaluate a rapid test introduced in the Federal District (FD).Objectives
To estimate the accuracy and reliability of the SD Bioeasy Dengue Duo rapid test and its components to detect dengue infections in a consecutive sample of symptomatic residents in the FD, Brazil.Methods
In total, 1353 venous blood samples were collected between 2013 and 2014. Two hundred and six positive samples (cases) and 246 negative samples (non cases) were required for sensitivity and specificity estimation, respectively; for agreement evaluation, we used 401 samples. The reference standard used was a composite of MAC-ELISA, virus isolation and real-time polymerase chain reaction (RT-qPCR). The evaluation was conducted prospectively under field conditions in the public health units of the FD.Findings
The results for the overall accuracy of the rapid test (NS1/IgM combined) showed 76% sensitivity and 98% specificity. The sensitivity for the NS1 component (67%) was better than that for the IgM component (35%). The positive likelihood ratio was 46, and the negative likelihood ratio was 0.24. The reliability of the test (NS1/IgM combined) demonstrated crude agreement of 98% (Kappa index 0.94).Main conclusions
The present phase III, large-scale validation study demonstrates that the rapid test SD Bioeasy Dengue Duo has moderate sensitivity (NS1/IgM combined) and high specificity. Therefore, the test is useful in confirming the diagnosis of dengue, but not enough to rule out the diagnosis. Our results also suggest that Dengue virus (DENV) viral load estimated through the RT-qPCR and antibody level measured through the MAC-ELISA could have had a direct influence on the accuracy of the rapid test.
SUBMITTER: Prado PS
PROVIDER: S-EPMC6014722 | biostudies-literature | 2018 Jun
REPOSITORIES: biostudies-literature
Memorias do Instituto Oswaldo Cruz 20180625 8
<h4>Background</h4>The diagnosis of dengue is complex. Until recently, only specialised laboratories were able to confirm dengue infection. However, this has changed with the newly available immunochromatographic rapid tests. Early diagnosis is of great interest, and point-of-care rapid tests have been increasingly used in Brazil. Most of those tests have not undergone validation in the Brazilian population. In this context, we decided to evaluate a rapid test introduced in the Federal District ...[more]