Field evaluation of quantitative point of care diagnostics to measure glucose-6-phosphate dehydrogenase activity.


ABSTRACT:

Background

Glucose-6-Phosphate dehydrogenase (G6PD) deficiency is the most common enzymopathy worldwide, no reliable bedside diagnostic tests to quantify G6PD activity exist. This study evaluated two novel quantitative G6PD diagnostics.

Methods

Participants with known G6PD activity were enrolled in Bangladesh. G6PD activity was measured by spectrophotometry, Biosensor (BS; AccessBio/CareStart, USA) and STANDARD G6PD (SG; SDBiosensor, ROK). G6PD activity was measured repeatedly in a subset of samples stored at room temperature and 4°C.

Results

158 participants were enrolled, 152 samples tested by BS, 108 samples by SG and 102 samples were tested by all three methods. In comparison to spectrophotometry BS had sensitivity and specificity of 72% (95%CI: 53-86) and 100% (95%CI: 97-100) at 30% cut off respectively, while SG had a sensitivity of 100% (95%CI: 88-100) and specificity of 97% (95%CI: 91-99) at the same cut off. The sensitivity and specificity at 70% cut off activity were 71% (95%CI: 59-82) and 98% (95%CI, 92-100) respectively for BS and 89% (95%CI: 77-96) and 93% (95%CI: 83-98) respectively for SG. When an optimal cut-off was applied the sensitivity of the BS at 70 cut off rose to 91% [95%CI: 80-96] and specificity to 82% [95%CI: 83-89]; a diagnostic accuracy comparable to that of the SG (p = 0.879). G6PD activity dropped significantly (-0.31U/gHb, 95%CI: -0.61 to -0.01, p = 0.022) within 24 hours in samples stored at room temperature, but did not fall below 90% of baseline activity until day 13 (-0.87U/gHb, 95%CI: (-1.11 to -0.62), p<0.001).

Conclusion

BS and SG are the first quantitative diagnostics to measure G6PD activity reliably at the bedside and represent suitable alternatives to spectrophotometry in resource poor settings. If samples are stored at 4°C, G6PD activity can be measured reliably for at least 7 days after sample collection.

SUBMITTER: Alam MS 

PROVIDER: S-EPMC6214512 | biostudies-literature | 2018

REPOSITORIES: biostudies-literature

altmetric image

Publications

Field evaluation of quantitative point of care diagnostics to measure glucose-6-phosphate dehydrogenase activity.

Alam Mohammad Shafiul MS   Kibria Mohammad Golam MG   Jahan Nusrat N   Thriemer Kamala K   Hossain Mohammad Sharif MS   Douglas Nicholas M NM   Phru Ching Swe CS   Khan Wasif Ali WA   Price Ric N RN   Ley Benedikt B  

PloS one 20181102 11


<h4>Background</h4>Glucose-6-Phosphate dehydrogenase (G6PD) deficiency is the most common enzymopathy worldwide, no reliable bedside diagnostic tests to quantify G6PD activity exist. This study evaluated two novel quantitative G6PD diagnostics.<h4>Methods</h4>Participants with known G6PD activity were enrolled in Bangladesh. G6PD activity was measured by spectrophotometry, Biosensor (BS; AccessBio/CareStart, USA) and STANDARD G6PD (SG; SDBiosensor, ROK). G6PD activity was measured repeatedly in  ...[more]

Similar Datasets

| S-EPMC8384181 | biostudies-literature
| S-EPMC6924682 | biostudies-literature
| S-EPMC6335905 | biostudies-literature
| S-EPMC7851131 | biostudies-literature
| S-EPMC283614 | biostudies-literature
| S-EPMC6595069 | biostudies-literature
| S-EPMC10882209 | biostudies-literature
| S-EPMC6821925 | biostudies-literature
| S-EPMC4365985 | biostudies-literature
| S-EPMC5488172 | biostudies-literature