Assessment of the modulation degrees of intensity-modulated radiation therapy plans.


ABSTRACT: BACKGROUND:To evaluate the modulation indices (MIs) for predicting the plan delivery accuracies of intensity-modulated radiation therapy (IMRT) plans. METHODS:A total of 100 dynamic IMRT plans that used TrueBeam STx and 102 dynamic IMRT plans that used Trilogy were selected. For each plan, various MIs were calculated, which included the modulation complexity score (MCS), plan-averaged beam area (PA), plan-averaged beam irregularity (PI), plan-averaged beam modulation (PM), MI quantifying multi-leaf collimator (MLC) speeds (MIs), MI quantifying MLC acceleration (MIa), and MI quantifying MLC acceleration and segment aperture irregularity (MIc,IMRT). To determine plan delivery accuracy, global gamma passing rates, MLC errors of log files, and dose-volumetric parameter differences between original and log file-reconstructed IMRT plans were obtained. To assess the ability of each MI for predicting plan delivery accuracy, Spearman's rank correlation coefficients (rs) between MIs and plan delivery accuracy measures were calculated. RESULTS:PI showed moderately strong correlations with gamma passing rates in MapCHECK2 measurements of both TrueBeam STx and Trilogy (rs?=?-?0.591 with p?

SUBMITTER: Park SY 

PROVIDER: S-EPMC6293636 | biostudies-literature | 2018 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Assessment of the modulation degrees of intensity-modulated radiation therapy plans.

Park So-Yeon SY   Kim Jung-In JI   Chun Minsoo M   Ahn Hyunjun H   Park Jong Min JM  

Radiation oncology (London, England) 20181213 1


<h4>Background</h4>To evaluate the modulation indices (MIs) for predicting the plan delivery accuracies of intensity-modulated radiation therapy (IMRT) plans.<h4>Methods</h4>A total of 100 dynamic IMRT plans that used TrueBeam STx and 102 dynamic IMRT plans that used Trilogy were selected. For each plan, various MIs were calculated, which included the modulation complexity score (MCS), plan-averaged beam area (PA), plan-averaged beam irregularity (PI), plan-averaged beam modulation (PM), MI quan  ...[more]

Similar Datasets

| S-EPMC8037748 | biostudies-literature
| S-EPMC5133459 | biostudies-literature
| S-EPMC7005113 | biostudies-literature
| S-EPMC5409532 | biostudies-literature
| S-EPMC3986382 | biostudies-other
| S-EPMC7449079 | biostudies-literature
| S-EPMC8655408 | biostudies-literature
| S-EPMC7557123 | biostudies-literature
| S-EPMC8214161 | biostudies-literature
| S-EPMC3903584 | biostudies-literature