Assessment of PCXMC for patients with different body size in chest and abdominal x ray examinations: a Monte Carlo simulation study.


ABSTRACT: A PC Program for x ray Monte Carlo (PCXMC) has been used to calculate organ doses in patient dosimetry and for the exposure assessment in epidemiological studies of radiogenic health related risks. This study compared the dosimetry from using the built-in stylized phantoms in the PCXMC to that of a newer hybrid phantom library with improved anatomical realism. We simulated chest and abdominal x ray projections for 146 unique body size computational phantoms, 77 males and 69 females, with different combinations of height (125-180?cm) and weight (20-140?kg) using the built-in stylized phantoms in the PCXMC version 2.0.1.4 and the hybrid phantom library using the Monte Carlo N-particle eXtended transport code 2.7 (MCNPX). Unfortunately, it was not possible to incorporate the hybrid phantom library into the PCXMC. We compared 14 organ doses, including dose to the active bone marrow, to evaluate differences between the built-in stylized phantoms in the PCXMC and the hybrid phantoms (Cristy and Eckerman 1987 Technical Report ORNL/TM-8381/V1, Oak Ridge National Laboratory, Eckerman and Ryman 1993 Technical Report 12 Oak Ridge, TN, Geyer et al 2014 Phys. Med. Biol. 59 5225-42). On average, organ doses calculated using the built-in stylized phantoms in the PCXMC were greater when compared to the hybrid phantoms. This is most prominent in AP abdominal exams by an average factor of 2.4-, 2.8-, and 2.8-fold for the 10-year-old, 15-year-old, and adult phantoms, respectively. For chest exams, organ doses are greater by an average factor of 1.1-, 1.4-, and 1.2-fold for the 10-year-old, 15-year-old, and adult phantoms, respectively. The PCXMX, due to its ease of use, is often selected to support dosimetry in epidemiological studies; however, it uses simplified models of the human anatomy that fail to account for variations in body morphometry for increasing weight. For epidemiological studies that use PCXMC dosimetry, associations between radiation-related disease risks and organ doses may be underestimated, and to a greater degree in pediatric, especially obese pediatric, compared to adult patients.

SUBMITTER: Borrego D 

PROVIDER: S-EPMC6376487 | biostudies-literature | 2018 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Assessment of PCXMC for patients with different body size in chest and abdominal x ray examinations: a Monte Carlo simulation study.

Borrego David D   Lowe Erin M EM   Kitahara Cari M CM   Lee Choonsik C  

Physics in medicine and biology 20180321 6


A PC Program for x ray Monte Carlo (PCXMC) has been used to calculate organ doses in patient dosimetry and for the exposure assessment in epidemiological studies of radiogenic health related risks. This study compared the dosimetry from using the built-in stylized phantoms in the PCXMC to that of a newer hybrid phantom library with improved anatomical realism. We simulated chest and abdominal x ray projections for 146 unique body size computational phantoms, 77 males and 69 females, with differe  ...[more]

Similar Datasets

| S-EPMC4999523 | biostudies-other
| S-EPMC9263979 | biostudies-literature
| S-EPMC4535364 | biostudies-literature
| S-EPMC3331834 | biostudies-other
| S-EPMC3978362 | biostudies-other
| S-EPMC2749500 | biostudies-literature
| S-EPMC7143414 | biostudies-literature
| S-EPMC7251717 | biostudies-literature
| S-EPMC2599820 | biostudies-literature