Comparative performance evaluation of four commercial multiplex real-time PCR assays for the detection of the diarrhoea-causing protozoa Cryptosporidium hominis/parvum, Giardia duodenalis and Entamoeba histolytica.


ABSTRACT:

Background

Multiplex molecular panels are relentlessly replacing conventional methods for the detection of enteric pathogens from stool samples in clinical and research laboratories. Here we evaluated four commercial multiplex real-time PCR assays for the detection of Cryptosporidium hominis/parvum, Giardia duodenalis and Entamoeba histolytica.

Methods

The diagnostic performance of the Gastroenteritis/Parasite Panel I (Diagenode), the RIDAGENE Parasitic Stool Panel (R-Biopharm), the Allplex Gastrointestinal Parasite Panel 4 (Seegene) and the FTD Stool Parasites (Fast Track) real-time PCR methods was assessed against a reference panel of 126 well-characterized DNA samples including Cryptosporidium hominis (n = 29), Cryptosporidium parvum (n = 3), Giardia duodenalis (n = 47), Entamoeba histolytica (n = 3), other parasite species (n = 20), and apparently healthy subjects (n = 24).

Principal findings

Obtained diagnostic sensitivities ranged from 53-88% for Cryptosporidium hominis/parvum, and from 68-100% for G. duodenalis. The R-Biopharm method achieved the best performance for the detection of Cryptosporidium hominis/parvum both in terms of diagnostic sensitivity (87.5%) and detection limit (a 100-fold increase compared to other tests). The Fast Track method was particularly suited for the detection of G. duodenalis, achieving a 100% sensitivity and a detection limit at least 10-fold superior. Detection of E. histolytica was similarly achieved by all compared methods except Diagenode.

Conclusions

Diagnostic performance varied largely depending on the method used and the targeted pathogen species. Factors including test sensitivity/specificity, cost, patient population surveyed, laboratory workflow, and diagnostic algorithm should be carefully considered when choosing the most appropriate multiplex PCR platform.

SUBMITTER: Paulos S 

PROVIDER: S-EPMC6453453 | biostudies-literature | 2019

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparative performance evaluation of four commercial multiplex real-time PCR assays for the detection of the diarrhoea-causing protozoa Cryptosporidium hominis/parvum, Giardia duodenalis and Entamoeba histolytica.

Paulos Silvia S   Saugar José María JM   de Lucio Aida A   Fuentes Isabel I   Mateo María M   Carmena David D  

PloS one 20190408 4


<h4>Background</h4>Multiplex molecular panels are relentlessly replacing conventional methods for the detection of enteric pathogens from stool samples in clinical and research laboratories. Here we evaluated four commercial multiplex real-time PCR assays for the detection of Cryptosporidium hominis/parvum, Giardia duodenalis and Entamoeba histolytica.<h4>Methods</h4>The diagnostic performance of the Gastroenteritis/Parasite Panel I (Diagenode), the RIDAGENE Parasitic Stool Panel (R-Biopharm), t  ...[more]

Similar Datasets

| S-EPMC446233 | biostudies-literature
| S-EPMC4599730 | biostudies-literature
| S-EPMC2413048 | biostudies-literature
| S-EPMC356880 | biostudies-literature
| S-EPMC2783587 | biostudies-literature
| S-EPMC3886425 | biostudies-other
| S-EPMC6722286 | biostudies-literature
| S-EPMC7767443 | biostudies-literature
| S-EPMC6256193 | biostudies-literature
| S-EPMC6631616 | biostudies-literature