Evaluation of the Precision Xtra meter for monitoring blood ?-hydroxybutyrate concentrations in late-gestation ewes.


ABSTRACT: Blood samples were collected from late-gestation ewes to determine the agreement of a point-of-care (POC) Precision Xtra meter and a standard laboratory test for ?-hydroxybutyrate (BHBA). Fresh whole blood samples were immediately tested with the POC instrument, and serum samples were analyzed with a standard commercial biochemical analyzer. Ewes were classified as having ketonemia if their BHBA concentrations were ?800?µmol/L. Scatter plots, paired t-tests, Bland-Altman limits of agreement, and Gwet AC1 tests were used to compare results. The 2 tests had very good agreement. The values between instruments were not statistically different based on paired t-tests ( p = 0.312). The intercept and slope of a linear mixed model, containing the standard test results as an outcome and the POC meter results as a predictor, were 0.02 (95% CI: 0.00, 0.04) and 0.98 (95% CI: 0.96, 1.01), respectively. When the samples were classified into ketonemic classes (non-ketonemic and ketonemic) based on BHBA concentrations obtained from each test, the Gwet AC1 statistic was 0.94 (95% CI: 0.91, 0.97; p < 0.001). The ketosis classification agreed in 95% of samples. Based on the Bland-Altman plot and limits of agreement, the optimal cutoff to diagnose ketonemia with the POC meter was 1,000?µmol/L, which is 200?µmol/L higher than the laboratory BHBA medical decision limit. The Precision Xtra meter provided excellent correlation and substantial agreement with the standard laboratory technique for measuring blood BHBA in late-gestation ewes.

SUBMITTER: Ratanapob N 

PROVIDER: S-EPMC6505764 | biostudies-literature | 2019 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of the Precision Xtra meter for monitoring blood β-hydroxybutyrate concentrations in late-gestation ewes.

Ratanapob Niorn N   VanLeeuwen John J   McKenna Shawn S   Wichtel Maureen M   Rodriguez-Lecompte Juan C JC   Menzies Paula P   Wichtel Jeffrey J  

Journal of veterinary diagnostic investigation : official publication of the American Association of Veterinary Laboratory Diagnosticians, Inc 20181212 1


Blood samples were collected from late-gestation ewes to determine the agreement of a point-of-care (POC) Precision Xtra meter and a standard laboratory test for β-hydroxybutyrate (BHBA). Fresh whole blood samples were immediately tested with the POC instrument, and serum samples were analyzed with a standard commercial biochemical analyzer. Ewes were classified as having ketonemia if their BHBA concentrations were ≥800 µmol/L. Scatter plots, paired t-tests, Bland-Altman limits of agreement, and  ...[more]

Similar Datasets

| S-EPMC7153552 | biostudies-literature
2022-07-18 | GSE195593 | GEO
| S-EPMC6316483 | biostudies-literature
| S-EPMC7807146 | biostudies-literature
| S-EPMC9448510 | biostudies-literature
| S-EPMC7591458 | biostudies-literature
| S-EPMC2913989 | biostudies-literature
| 2719536 | ecrin-mdr-crc
| S-EPMC6541830 | biostudies-literature
| S-EPMC7520093 | biostudies-literature