Evaluation of a duplex reverse-transcription real-time PCR assay for the detection of encephalomyocarditis virus.


ABSTRACT: We evaluated a fluorogenic probe-based assay for the detection of encephalomyocarditis virus (EMCV) by comparing a set of published primers and probe to a new set of primers and probe. The published reagents failed to amplify a range of Australian isolates and an Italian reference strain of EMCV. In contrast, an assay based on 2 new sets of primers and probes that were run in a duplex reverse-transcription real-time PCR (RT-rtPCR) worked well, with high amplification efficiency. The analytical sensitivity was ~100-fold higher than virus isolation in cell culture. The intra-assay variation was 0.21-4.90%. No cross-reactivity was observed with a range of other porcine viruses. One hundred and twenty-two clinical specimens were tested simultaneously by RT-rtPCR and virus isolation in cell culture; 72 specimens gave positive results by RT-rtPCR, and 63 of these were also positive by virus isolation. Of 245 archived cell culture isolates of EMCV that were tested in the RT-rtPCR, 242 samples were positive. The new duplex RT-rtPCR assay is a reliable tool for the detection of EMCV in clinical specimens and for use in epidemiologic investigations.

SUBMITTER: Qin S 

PROVIDER: S-EPMC6505904 | biostudies-literature | 2018 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of a duplex reverse-transcription real-time PCR assay for the detection of encephalomyocarditis virus.

Qin Shaomin S   Underwood Darren D   Driver Luke L   Kistler Carol C   Diallo Ibrahim I   Kirkland Peter D PD  

Journal of veterinary diagnostic investigation : official publication of the American Association of Veterinary Laboratory Diagnosticians, Inc 20180602 4


We evaluated a fluorogenic probe-based assay for the detection of encephalomyocarditis virus (EMCV) by comparing a set of published primers and probe to a new set of primers and probe. The published reagents failed to amplify a range of Australian isolates and an Italian reference strain of EMCV. In contrast, an assay based on 2 new sets of primers and probes that were run in a duplex reverse-transcription real-time PCR (RT-rtPCR) worked well, with high amplification efficiency. The analytical s  ...[more]

Similar Datasets

| S-EPMC2238069 | biostudies-literature
| S-EPMC4733162 | biostudies-literature
| S-EPMC6505849 | biostudies-literature
| S-EPMC2045251 | biostudies-literature
| S-EPMC522381 | biostudies-literature
| S-EPMC4710529 | biostudies-literature
| S-EPMC321536 | biostudies-literature
| S-EPMC2897487 | biostudies-literature
| S-EPMC4400749 | biostudies-literature
| S-EPMC3412361 | biostudies-literature