Unknown

Dataset Information

0

Calibration-Free Electrochemical Biosensors Supporting Accurate Molecular Measurements Directly in Undiluted Whole Blood.


ABSTRACT: The need to calibrate to correct for sensor-to-sensor fabrication variation and sensor drift has proven a significant hurdle in the widespread use of biosensors. To maintain clinically relevant (±20% for this application) accuracy, for example, commercial continuous glucose monitors require recalibration several times a day, decreasing convenience and increasing the chance of user errors. Here, however, we demonstrate a "dual-frequency" approach for achieving the calibration-free operation of electrochemical biosensors that generate an output by using square-wave voltammetry to monitor binding-induced changes in electron transfer kinetics. Specifically, we use the square-wave frequency dependence of their response to produce a ratiometric signal, the ratio of peak currents collected at responsive and non- (or low) responsive square-wave frequencies, which is largely insensitive to drift and sensor-to-sensor fabrication variations. Using electrochemical aptamer-based (E-AB) biosensors as our test bed, we demonstrate the accurate and precise operation of sensors against multiple drugs, achieving accuracy in the measurement of their targets of within better than 20% across dynamic ranges of up to 2 orders of magnitude without the need to calibrate each individual sensor.

SUBMITTER: Li H 

PROVIDER: S-EPMC6519131 | biostudies-literature | 2017 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Calibration-Free Electrochemical Biosensors Supporting Accurate Molecular Measurements Directly in Undiluted Whole Blood.

Li Hui H   Dauphin-Ducharme Philippe P   Ortega Gabriel G   Plaxco Kevin W KW  

Journal of the American Chemical Society 20170802 32


The need to calibrate to correct for sensor-to-sensor fabrication variation and sensor drift has proven a significant hurdle in the widespread use of biosensors. To maintain clinically relevant (±20% for this application) accuracy, for example, commercial continuous glucose monitors require recalibration several times a day, decreasing convenience and increasing the chance of user errors. Here, however, we demonstrate a "dual-frequency" approach for achieving the calibration-free operation of el  ...[more]

Similar Datasets

| S-EPMC4892350 | biostudies-literature
| S-EPMC6792296 | biostudies-literature
| S-EPMC5318951 | biostudies-literature
| S-EPMC3082468 | biostudies-literature
| S-EPMC6649092 | biostudies-literature
| S-EPMC7152911 | biostudies-literature
| S-EPMC8132996 | biostudies-literature
| S-EPMC4647221 | biostudies-literature