A Continuous Flow-through Microfluidic Device for Electrical Lysis of Cells.


ABSTRACT: In contrast to the delicate 3D electrodes in the literature, a simple flow-through device is proposed here for continuous and massive lysis of cells using electricity. The device is essentially a rectangular microchannel with a planar electrode array built on its bottom wall, actuated by alternating current (AC) voltages between neighboring electrodes, and can be incorporated easily into other biomedical systems. Human whole blood diluted 10 times with phosphate-buffered saline (about 6 108 cells per mL) was pumped through the device, and the cells were completely lysed within 7 s after the application of a 20 V peak-to-peak voltage at 1 MHz, up to 400 ?L/hr. Electric field and Maxwell stress were calculated for assessing electrical lysis. Only the lower half-channel was exposed to an electric field exceeding the irreversible threshold value of cell electroporation (Eth2), suggesting that a cross flow, proposed here primarily as the electro-thermally induced flow, was responsible for bringing the cells in the upper half-channel downward to the lower half-channel. The Maxwell shear stress associated with Eth2 was one order of magnitude less than the threshold mechanical stresses for lysis, implying that an applied moderate mechanical stress could aid electrical lysis.

SUBMITTER: Lo YJ 

PROVIDER: S-EPMC6523307 | biostudies-literature | 2019 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

A Continuous Flow-through Microfluidic Device for Electrical Lysis of Cells.

Lo Ying-Jie YJ   Lei U U  

Micromachines 20190413 4


In contrast to the delicate 3D electrodes in the literature, a simple flow-through device is proposed here for continuous and massive lysis of cells using electricity. The device is essentially a rectangular microchannel with a planar electrode array built on its bottom wall, actuated by alternating current (AC) voltages between neighboring electrodes, and can be incorporated easily into other biomedical systems. Human whole blood diluted 10 times with phosphate-buffered saline (about 6 10<sup>8  ...[more]

Similar Datasets

| S-EPMC2923460 | biostudies-literature
| S-EPMC4462155 | biostudies-literature
| S-EPMC6406949 | biostudies-literature
| S-EPMC7245095 | biostudies-literature
| S-EPMC4626017 | biostudies-literature
| S-EPMC6591506 | biostudies-literature
2015-07-21 | GSE65516 | GEO
| S-EPMC4558391 | biostudies-literature
| S-EPMC5103028 | biostudies-literature
| S-EPMC5744346 | biostudies-literature