Unknown

Dataset Information

0

Cerebrospinal fluid-suppressed T2 -weighted MR imaging at 7 T for human brain.


ABSTRACT:

Purpose

T2 -weighted lesional imaging is most commonly performed using inversion recovery turbo spin echoes. At 7 T, however, this acquisition is limited for specific absorption rate and resolution. This work describes and implements a method to generate CSF-suppressed T2 -weighted imaging.

Methods

The strategy uses a driven equilibrium spin-echo preparation within an inversion recovery with multiple 3D gradient-echo imaging blocks. Images are combined using the self-normalization approach, which achieves CSF suppression through optimized timing of individual blocks and minimizes sources of variation due to coil receptivity, T2 * , and proton density. Simulations of the magnetization-prepared fluid-attenuated inversion recovery gradient-echo (MPFLAGRE) method over T1 and T2 relaxation values are performed, and in vivo demonstrations using an 8 × 2 transceiver array in healthy controls are shown.

Results

The specific absorption rate of the calculated MPFLAGRE sequence is 11.1 ± 0.5 W (n = 5 volunteers), which is 74 ± 2% of the US Food and Drug Administration guidelines. This method acquires both contrasts for CSF suppression with detection of long T2 components and T2 -weighted imaging in a single acquisition. In healthy controls, the former contrast generates increased signal in the cortical rim and ependyma. A comparison is shown with a conventional 3D SPACE fluid-attenuated inversion recovery acquisition, and sensitivity to pathology is demonstrated in an epilepsy patient.

Conclusion

As applied with the 8 × 2 transceiver, the MPFLAGRE sequence generates both whole-brain contrast suitable for lesional and T2 -weighted imaging at 7 T in fewer than 10 minutes within the US Food and Drug Administration's specific absorption rate guidelines.

SUBMITTER: Pan JW 

PROVIDER: S-EPMC6590483 | biostudies-literature | 2019 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Cerebrospinal fluid-suppressed T<sub>2</sub> -weighted MR imaging at 7 T for human brain.

Pan Jullie W JW   Moon Chan Hong CH   Hetherington Hoby P HP  

Magnetic resonance in medicine 20181119 5


<h4>Purpose</h4>T<sub>2</sub> -weighted lesional imaging is most commonly performed using inversion recovery turbo spin echoes. At 7 T, however, this acquisition is limited for specific absorption rate and resolution. This work describes and implements a method to generate CSF-suppressed T<sub>2</sub> -weighted imaging.<h4>Methods</h4>The strategy uses a driven equilibrium spin-echo preparation within an inversion recovery with multiple 3D gradient-echo imaging blocks. Images are combined using  ...[more]

Similar Datasets

| S-EPMC8267868 | biostudies-literature
| S-EPMC8113983 | biostudies-literature
| S-EPMC9434732 | biostudies-literature
| S-EPMC8050458 | biostudies-literature
| S-EPMC9113559 | biostudies-literature
| S-EPMC7787526 | biostudies-literature
| S-EPMC7964624 | biostudies-literature
| S-EPMC9730674 | biostudies-literature
| S-EPMC8356990 | biostudies-literature
| PRJNA856347 | ENA