Clinical Applicability of a Textile 1-Lead ECG Device for Overnight Monitoring.


ABSTRACT: Even for 1-lead electrocardiography (ECG), single-use gel conductive electrodes are employed in a clinical setting. However, gel electrodes show limited applicability for long-term monitoring due to skin irritation and detachment. In the present study, we investigated the validity of a textile ECG-belt suitable for long-term measurements in clinical use. In order to assess the signal quality and validity of the ECG-belt during sleep, 242 patients (186 males and 56 females, age 52 (interquartile range 42-60) years, body mass index 29 (interquartile range 26-33) kg·m-2) with suspected sleep apnoea underwent overnight polysomnography including standard 1-lead ECG recording. The single intervals between R-peaks (RR-intervals) were calculated from the ECG-signals. We found a mean difference for average RR-intervals of -2.9 ms, a standard error of estimate of 0.39%, as well as a Pearson r of 0.91. Furthermore, we found that the validity of the ECG-belt decreases when lying on the side, which was potentially due to the fitting of the belt. In conclusion, the validity of RR-interval measurements using the ECG-belt is high and it may be further improved for future applications by optimizing wear fitting.

SUBMITTER: Fontana P 

PROVIDER: S-EPMC6603515 | biostudies-literature | 2019 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Clinical Applicability of a Textile 1-Lead ECG Device for Overnight Monitoring.

Fontana Piero P   Martins Neusa R Adão NRA   Camenzind Martin M   Rossi René M RM   Baty Florent F   Boesch Maximilian M   Schoch Otto D OD   Brutsche Martin H MH   Annaheim Simon S  

Sensors (Basel, Switzerland) 20190528 11


Even for 1-lead electrocardiography (ECG), single-use gel conductive electrodes are employed in a clinical setting. However, gel electrodes show limited applicability for long-term monitoring due to skin irritation and detachment. In the present study, we investigated the validity of a textile ECG-belt suitable for long-term measurements in clinical use. In order to assess the signal quality and validity of the ECG-belt during sleep, 242 patients (186 males and 56 females, age 52 (interquartile  ...[more]

Similar Datasets

| S-EPMC6696177 | biostudies-literature
| S-EPMC7669461 | biostudies-literature
| S-EPMC7109157 | biostudies-literature
| S-EPMC7050507 | biostudies-literature
| S-EPMC7917757 | biostudies-literature
| S-EPMC10648477 | biostudies-literature
| S-EPMC7296680 | biostudies-literature
| S-EPMC8195254 | biostudies-literature
| S-EPMC7581708 | biostudies-literature
| S-EPMC7180963 | biostudies-literature