How have advances in CT dosimetry software impacted estimates of CT radiation dose and cancer incidence? A comparison of CT dosimetry software: Implications for past and future research.


ABSTRACT:

Objective

Organ radiation dose from a CT scan, calculated by CT dosimetry software, can be combined with cancer risk data to estimate cancer incidence resulting from CT exposure. We aim to determine to what extent the use of improved anatomical representation of the adult human body "phantom" in CT dosimetry software impacts estimates of radiation dose and cancer incidence, to inform comparison of past and future research.

Methods

We collected 20 adult cases for each of three CT protocols (abdomen/pelvis, chest and head) from each of five public hospitals (random sample) (January-April inclusive 2010) and three private clinics (self-report). Organ equivalent and effective dose were calculated using both ImPACT (mathematical phantom) and NCICT (voxelised phantom) software. Bl