ABSTRACT: Background
An External Quality Assessment (EQA) program was developed to investigate the status of estrogen receptor (ER), progesterone receptor (PR), and Ki-67 immunohistochemical (IHC) detection in breast cancer and to evaluate the reproducibility of staining and interpretation in 44 pathology laboratories in China.Methods
This program was implemented through three specific steps. In study I, three revising centres defined the reference value for 11 sections. In study II, 41 participating centres (PC) stained and interpreted 11 sections by their own daily practice IHC protocols. In study III, all cases received second interpretation opinions.Results
The stained slides of 44 laboratories were up to the interpretation standard. The overall interpretation concordance rate of this study was over 90%. A perfect agreement was reached among the PCs for the cases with ER+ and PR+ >?50% and Ki-67?>?30%, whereas a moderate agreement was observed for intermediate categories. After second interpretations, the misclassification rates for ER were reduced by 12.20%, for PR were reduced by 17.07%, and for Ki-67 were reduced by 4.88%. Up to 31 PCs observed a benefit from the second opinion strategy.Conclusions
This project is the first EQA study performed on a national scale for assessment of ER, PR and Ki-67 status by IHC in China. In the whole IHC evaluation process, the intermediate categories were less reproducible than those with high expression rates. Second opinions can significantly improve the diagnostic agreement of pathologists' interpretations.
SUBMITTER: Pu T
PROVIDER: S-EPMC6805628 | biostudies-literature | 2019 Oct
REPOSITORIES: biostudies-literature
Pu Tianjie T Shui Ruohong R Shi Jie J Liang Zhiyong Z Yang Wentao W Bu Hong H Li Qin Q Zhang Zhang Z
BMC cancer 20191022 1
<h4>Background</h4>An External Quality Assessment (EQA) program was developed to investigate the status of estrogen receptor (ER), progesterone receptor (PR), and Ki-67 immunohistochemical (IHC) detection in breast cancer and to evaluate the reproducibility of staining and interpretation in 44 pathology laboratories in China.<h4>Methods</h4>This program was implemented through three specific steps. In study I, three revising centres defined the reference value for 11 sections. In study II, 41 pa ...[more]