Feasibility Studies of an Insulin-Only Bionic Pancreas in a Home-Use Setting.


ABSTRACT: BACKGROUND:We tested the safety and performance of the "insulin-only" configuration of the bionic pancreas (BP) closed-loop blood-glucose control system in a home-use setting to assess glycemic outcomes using different static and dynamic glucose set-points. METHOD:This is an open-label non-randomized study with three consecutive intervention periods. Participants had consecutive weeks of usual care followed by the insulin-only BP with (1) an individualized static set-point of 115 or 130 mg/dL and (2) a dynamic set-point that automatically varied within 110 to 130 mg/dL, depending on hypoglycemic risk. Human factors (HF) testing was conducted using validated surveys. The last five days of each study arm were used for data analysis. RESULTS:Thirteen participants were enrolled with a mean age of 28 years, mean A1c of 7.2%, and mean daily insulin dose of 0.6 U/kg (0.4-1.0 U/kg). The usual care arm had an average glucose of 145 ± 20 mg/dL, which increased in the static set-point arm (159 ± 8 mg/dL, P = .004) but not in the dynamic set-point arm (154 ± 10 mg/dL, P = ns). There was no significant difference in time spent in range (70-180 mg/dL) among the three study arms. There was less time <70 mg/dL with both the static (1.8% ± 1.4%, P = .009) and dynamic set-point (2.7±1.5, P = .051) arms compared to the usual-care arm (5.5% ± 4.2%). HF testing demonstrated preliminary user satisfaction and no increased risk of diabetes burden or distress. CONCLUSIONS:The insulin-only configuration of the BP using either static or dynamic set-points and initialized only with body weight performed similarly to other published insulin-only systems.

SUBMITTER: Ekhlaspour L 

PROVIDER: S-EPMC6835195 | biostudies-literature | 2019 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications


<h4>Background</h4>We tested the safety and performance of the "insulin-only" configuration of the bionic pancreas (BP) closed-loop blood-glucose control system in a home-use setting to assess glycemic outcomes using different static and dynamic glucose set-points.<h4>Method</h4>This is an open-label non-randomized study with three consecutive intervention periods. Participants had consecutive weeks of usual care followed by the insulin-only BP with (1) an individualized static set-point of 115  ...[more]

Similar Datasets

| S-EPMC5358809 | biostudies-literature
| S-EPMC8266971 | biostudies-literature
| S-EPMC9634987 | biostudies-literature
| S-EPMC9529301 | biostudies-literature
| S-EPMC4183762 | biostudies-literature
| S-EPMC10028490 | biostudies-literature
| S-EPMC7042391 | biostudies-literature
| S-EPMC3476884 | biostudies-literature
| S-EPMC3687268 | biostudies-literature