ABSTRACT: Purpose
The robust and reliable utilization of proton magnetic resonance spectroscopic imaging (MRSI) at high fields is hampered by several key technical difficulties, including contamination from extracranial lipids. To that end, this work presents novel lipid suppression sequences for proton MRSI in the human brain utilizing elliptical localization with pulsed second-order fields (ECLIPSE).Methods
Two lipid suppression methods were implemented with the ECLIPSE gradient insert. One method is a variable power, 4-pulse sequence optimized to achieve outer volume suppression (OVS) and compared against a standard, 8-slice OVS method. The second ECLIPSE method is implemented as an inversion recovery (IR) sequence with elliptical inner volume selection (IVS) and compared against a global IR method.Results
The ECLIPSE-OVS sequence provided a 116-fold mean lipid suppression (range, 104-134), whereas an optimized 8-slice OVS sequence achieved 15-fold suppression (range, 13-18). Furthermore, the superior ECLIPSE-OVS suppression was achieved at 30% of the radiofrequency (RF) power required by 8-slice OVS. The ECLIPSE-based IR sequence suppressed skull lipids by 155-fold (range, 122-257), compared to 16-fold suppression (range, 14-19) achieved with IR.Conclusion
OVS and IVS executed with ECLIPSE provide robust and effective lipid suppression at reduced RF power with high immunity to variations in B1 and T1 .