Development of a Standardized Chemiluminescence Immunoassay for the Detection of Autoantibodies Against Human M-Type Phospholipase A2 Receptor in Primary Membranous Nephropathy.
Ontology highlight
ABSTRACT: Introduction:Autoantibodies against the M-type phospholipase A2 receptor (PLA2R) are important markers in the diagnosis and monitoring of primary membranous nephropathy (pMN). For the detection of anti-PLA2R autoantibodies, a standardized recombinant cell-based indirect immunofluorescence assay (RC-IFA) and enzyme-linked immunosorbent assay (ELISA) are widely used, the former providing higher sensitivity but lacking a finely graduated quantification of antibody titers. In this study, we evaluated the diagnostic performance characteristics of a novel standardized chemiluminescence immunoassay (ChLIA) by comparison with the established anti-PLA2R test systems. Methods:Sera from 155 patients with biopsy-proven pMN and 154 disease controls were analyzed for autoantibodies against PLA2R by the novel ChLIA as well as by ELISA and RC-IFA. Results:The clinical sensitivity of the ChLIA (83.9%) was higher compared with ELISA (73.5%) and equaled that of RC-IFA (83.2%), at similar specificities (?99.4%). Among ELISA-negative pMN samples, ChLIA and RC-IFA yielded positive results in 39.0% and 36.6%, respectively. The qualitative agreement amounted to 94.5% (ChLIA vs. ELISA) and 99.4% (ChLIA vs. RC-IFA). Conclusion:The novel anti-PLA2R ChLIA outperforms the ELISA in detecting patients with pMN and demonstrates almost perfect agreement with RC-IFA. It thus presents a promising alternative tool for accurate anti-PLA2R testing, with the advantage of rapid turnaround times and fully automated random-access processing.
SUBMITTER: Dahnrich C
PROVIDER: S-EPMC7000842 | biostudies-literature | 2020 Feb
REPOSITORIES: biostudies-literature
ACCESS DATA