Ontology highlight
ABSTRACT: Background
The detection of viral respiratory tract infections has evolved greatly with the development of PCR based commercial systems capable of simultaneously detecting a wide variety of pathogens.Objectives
Evaluate the relative performance of two commercial broad range systems for the detection of viral agents in clinical respiratory tract specimens from immunocompromised children.Study design
A total of 176 patient samples were included in the analysis, representing only the first sample collected for each patient, and excluding failed reactions. Samples were de-identified and assayed in parallel using two different, broadly multiplexed PCR systems: ResPlex™ II Panel v2.0 (ResPlex), Qiagen, Hilden, Germany and FilmArray(®) Respiratory Panel (FilmArray), Idaho Technology Inc., Salt Lake City, UT. Method comparison was based upon pair-wise concordance of results according to patient age, viral target and number of targets detected.Results
The two systems showed an overall concordance, by patient, of 83.8% (p=0.0001). FilmArray detected at least one target in 68.8% of samples, while ResPlex detected at least one target in 56.8%. ResPlex failed to detect 20.7% of FilmArray positives, and FilmArray failed to detect 4% of ResPlex positives. The relative performance of each system (including which system detected a higher number of positive samples) varied when stratified by target viral pathogen.Conclusions
Broadly multiplexed PCR is an effective means of detecting large numbers of clinically relevant respiratory viral pathogens.
SUBMITTER: Hayden RT
PROVIDER: S-EPMC7108354 | biostudies-literature | 2012 Apr
REPOSITORIES: biostudies-literature
Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology 20120131 4
<h4>Background</h4>The detection of viral respiratory tract infections has evolved greatly with the development of PCR based commercial systems capable of simultaneously detecting a wide variety of pathogens.<h4>Objectives</h4>Evaluate the relative performance of two commercial broad range systems for the detection of viral agents in clinical respiratory tract specimens from immunocompromised children.<h4>Study design</h4>A total of 176 patient samples were included in the analysis, representing ...[more]