A simple and compact smartphone-based device for the quantitative readout of colloidal gold lateral flow immunoassay strips.


ABSTRACT: Colloidal gold lateral flow immunoassay strips (AuNPs-LFIS) have been widely applied as qualitative diagnostic tools for point-of-care tests (POCT). If strip readers were incorporated, their use could be extended to quantitative analysis. However, their cost and non-portability render commercial strip readers unavailable for use in either home testing, community or rural hospital diagnosis. This is particularly true for on-site testing. Here, a smartphone-based reader was designed and 3D-printed for quantitatively assess AuNPs-LFIS. The basic principle of the devise was relying on a smartphone's ambient light sensor (SPALS). This sensor was harnessed to measure the transmitted light intensities originating from the T-lines on the strips, the transmitted light intensities vary with concentration of AuNP on the T-lines. To validate this approach, our newly developed smartphone's ambient light sensor-based reader (SPALS-reader) was used to readout AuNPs-LFIS of three analytical targets: cadmium ion (Cd2+; limit of detection (LOD) was 0.16?ng/mL), clenbuterol (CL; LOD was 0.046?ng/mL), and porcine epidemic diarrhea virus (PEDV; LOD was 0.055??g/mL). The result showed good consistency with the results of conventional image analysis approaches, indicating that the smartphone-based device is appropriate for use in AuNPs-LFIS readouts. Compared with the traditional analysis method, the developed AuNPs-LFIS reader is easier operated, lower cost and more portable, which provided an on-site quantitative analysis tool for AuNPs-LFIS and enhances the applied range of AuNPs-LFIS.

SUBMITTER: Xiao W 

PROVIDER: S-EPMC7127147 | biostudies-literature | 2018 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

A simple and compact smartphone-based device for the quantitative readout of colloidal gold lateral flow immunoassay strips.

Xiao Wei W   Huang Caihong C   Xu Fei F   Yan Junjie J   Bian Hongfen H   Fu Qiangqiang Q   Xie Kaixin K   Wang Lei L   Tang Yong Y  

Sensors and actuators. B, Chemical 20180320


Colloidal gold lateral flow immunoassay strips (AuNPs-LFIS) have been widely applied as qualitative diagnostic tools for point-of-care tests (POCT). If strip readers were incorporated, their use could be extended to quantitative analysis. However, their cost and non-portability render commercial strip readers unavailable for use in either home testing, community or rural hospital diagnosis. This is particularly true for on-site testing. Here, a smartphone-based reader was designed and 3D-printed  ...[more]

Similar Datasets

| S-EPMC6339659 | biostudies-literature
| S-EPMC9098456 | biostudies-literature
| S-EPMC10018695 | biostudies-literature
| S-EPMC8869244 | biostudies-literature
| S-EPMC7418077 | biostudies-literature
| S-EPMC4113134 | biostudies-literature
| S-EPMC5703415 | biostudies-literature
| S-EPMC5430532 | biostudies-literature
| S-EPMC6988510 | biostudies-literature
| S-EPMC10017168 | biostudies-literature