Unknown

Dataset Information

0

Effective chemical virus inactivation of patient serum compatible with accurate serodiagnosis of infections.


ABSTRACT:

Objectives

Highly pathogenic viruses such as EBOV are a threat to routine laboratory workers. Inactivation procedures with Triton X-100 0.1% and/or heat are currently recommended, but have unknown effects on the accuracy of serological testing. Furthermore, virus inactivation by Triton X-100 0.1% was shown to be ineffective in serum. This study aimed to demonstrate virus inactivation in serum by Triton X-100 1% and maintained accuracy of serological testing.

Methods

A panel of 19 serological tests was run on patient serum samples after treatment with Triton X-100 1%, 0.1%, and 0.1% + heat inactivation at 60°C for 1 h. Mean differences between measurements (bias) were calculated applying the Bland-Altman method. To determine effectiveness of virus inactivation, herpes simplex virus 1 (HSV-1) was spiked into medium containing 90% or 1% serum, and treated with Triton X-100 0.1% or 1%. Infectious titres were then determined on Vero cells.

Results

Serological measurements showed good agreement between controls and samples treated with Triton X-100 0.1% and 1%, with an estimated bias of 0.6 ± 9.2% (n = 258) and -0.1 ± 18.6% (n = 174), respectively. Discordant qualitative results were rare. Conversely, heat inactivation alone and combined with Triton X-100 0.1% triggered a bias of 17.5 ± 66.4% (n = 200) and 37.9 ± 79.8% (n = 160), respectively. Triton X-100 1% completely inactivated HSV-1 in 1% and 90% serum while Triton X-100 0.1% failed to do so in 90% serum.

Conclusions

Unlike heat inactivation, Triton X-100 1% enabled accurate serological testing and completely inactivated HSV-1 in serum. This simple method could allow safe routine serological diagnostics in high-risk patients.

SUBMITTER: Remy MM 

PROVIDER: S-EPMC7128130 | biostudies-literature | 2019 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Effective chemical virus inactivation of patient serum compatible with accurate serodiagnosis of infections.

Remy M M MM   Alfter M M   Chiem M-N MN   Barbani M T MT   Engler O B OB   Suter-Riniker F F  

Clinical microbiology and infection : the official publication of the European Society of Clinical Microbiology and Infectious Diseases 20181028 7


<h4>Objectives</h4>Highly pathogenic viruses such as EBOV are a threat to routine laboratory workers. Inactivation procedures with Triton X-100 0.1% and/or heat are currently recommended, but have unknown effects on the accuracy of serological testing. Furthermore, virus inactivation by Triton X-100 0.1% was shown to be ineffective in serum. This study aimed to demonstrate virus inactivation in serum by Triton X-100 1% and maintained accuracy of serological testing.<h4>Methods</h4>A panel of 19  ...[more]

Similar Datasets

| S-EPMC4918181 | biostudies-literature
| S-EPMC7001000 | biostudies-literature
| S-EPMC7547523 | biostudies-literature
| S-EPMC2952994 | biostudies-literature
| S-EPMC4980329 | biostudies-literature
| S-EPMC6420165 | biostudies-literature
| S-EPMC6425182 | biostudies-literature
| S-EPMC9974437 | biostudies-literature
| S-EPMC4838304 | biostudies-literature
| S-EPMC6759188 | biostudies-literature