Outcomes based on blood pressure in patients on continuous flow left ventricular assist device support: An Interagency Registry for Mechanically Assisted Circulatory Support analysis.


ABSTRACT:

Background

An optimal blood pressure (BP) range to mitigate morbidity and mortality on left ventricular assist device (LVAD) support has not been clearly defined.

Methods

Average Doppler opening pressure, mean arterial pressure (MAP), and/or systolic blood pressure (SBP) were calculated in operative survivors (n = 16,155) of LVAD support in the Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS). BP distributions were used to group patients into low (BP <25th percentile), normal (25-75th percentile), high (75th-95th percentile), and very high (>95th percentile). Associations between BP and adverse events were evaluated using Cox regression (hazard ratio[HR], 95% confidence interval).

Results

The median (25th, 75th) MAP, Doppler, and SBP (mm Hg) during continuous flow LVAD support were 84 (77, 90), 85 (80, 92), and 99 (90, 107) mm Hg, respectively. BP had a bimodal risk association with survival. At 3 years, survival was 58% ± 1.8% in those with low MAP (≤75 mm Hg) vs 70% ± 0.9%, 71% ± 1.5%, and 63% ± 3.0% in the those with normal, high, or very high average MAP, respectively. Patients with chronically low MAP (≤75 mm Hg), Doppler (≤80 mm Hg), and SBP (<90 mm Hg) had 35%-42% higher adjusted hazards of death than patients with normal or high BP (p ≤ 0.0001). Patients with MAP >100 mm Hg, Doppler ≥105 mm Hg, and SBP ≥120 mm Hg had 17%-20% higher adjusted hazards of death than those with normal pressures (p < 0.05). In patients on axial flow LVADs, elevated SBP (HR 1.08 [95% confidence interval, 1.04-1.13] per 10 mm Hg increase) but not MAP correlated with increased incident of stroke.

Conclusions

In INTERMACS, BP extremes during LVAD support increase the risk for adverse events, supporting a MAP goal >75 mm Hg and <90 mm Hg. Hypotension conferred the highest risk for mortality. Excessive BP control should be avoided, and Doppler opening pressure should not be assumed to represent MAP in all patients.

SUBMITTER: Cowger JA 

PROVIDER: S-EPMC7200289 | biostudies-literature | 2020 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Outcomes based on blood pressure in patients on continuous flow left ventricular assist device support: An Interagency Registry for Mechanically Assisted Circulatory Support analysis.

Cowger Jennifer A JA   Shah Palak P   Pagani Francis D FD   Grafton Gillan G   Stulak John J   Chamogeorgakis Themistokles T   Lanfear David D   Nemeh Hassan H   Pinney Sean S  

The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation 20191126 5


<h4>Background</h4>An optimal blood pressure (BP) range to mitigate morbidity and mortality on left ventricular assist device (LVAD) support has not been clearly defined.<h4>Methods</h4>Average Doppler opening pressure, mean arterial pressure (MAP), and/or systolic blood pressure (SBP) were calculated in operative survivors (n = 16,155) of LVAD support in the Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS). BP distributions were used to group patients into low (BP  ...[more]

Similar Datasets

| S-EPMC8854346 | biostudies-literature
| S-EPMC5717751 | biostudies-literature
| S-EPMC4475666 | biostudies-literature
| S-EPMC4943678 | biostudies-literature
| S-EPMC7702162 | biostudies-literature
| S-EPMC6092931 | biostudies-literature
2003-09-01 | GSE430 | GEO
| S-EPMC6217859 | biostudies-literature
| S-EPMC3781792 | biostudies-literature
| S-EPMC7334284 | biostudies-literature