Evaluation of a novel portable three-dimensional scapular kinematics assessment system with inter and intraobserver reproducibility and normative data for healthy adults.


ABSTRACT: PURPOSE:To evaluate the intra and interobserver reproducibility of a new system that assesses the three-dimensional humero-scapulo-thoracic kinematics using wearable technology in an outpatient setting. To obtain normative data with the system for scapular angular motions in three planes. METHODS:The SHoW Motion 3D kinematic tracking system is a motion analysis system that uses wireless wearable non-invasive inertial-magnetic sensors to assess the three-dimensional kinematics of the shoulder girdle. The sensors are placed over the skin in the sternum, scapular spine and arm to precisely define angular motions of the humerus and the scapula with three Degrees of Freedom (DOF) for each segment. The system was used to measure the scapular angular motions in three planes (upward/downward rotation, internal/external rotation and anterior/posterior tilt) during two shoulder full-range movements (flexion/extension and abduction/abduction) in both shoulders of 25 healthy volunteers (13 males and 12 females, mean age: 37 [standard deviation 11.1] years). In a first measuring session one examiner made two evaluations alternating with another examiner that made a third evaluation. In a second session, one week apart, the first examiner made a fourth evaluation. A mean curve was computed from the normalized data for each measurement to obtain normative data for scapular angular kinematics. Intra and inter-observer reproducibility was evaluated using Root Mean Square Error Estimation (RMSE) and Coefficients for Multiple Correlations (CMC). RESULTS:Both shoulders of the 25 volunteers were evaluated four times. The two hundred resulting kinematic analyses were pooled to get normative values for relations between humeral elevation angles and the three angular movements of the scapula. The system showed at least very good (CMC?>?0.90) intra and inter-observer reproducibility for scapular tilt and upward-downward rotations both in flexion and abduction. For scapular internal-external rotation the results were acceptable (CMC?>?0.75) but not as good, especially for the abduction movement. RMSE calculations showed consistently good reproducibility with RSME

SUBMITTER: Ruiz Iban MA 

PROVIDER: S-EPMC7221052 | biostudies-literature | 2020 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of a novel portable three-dimensional scapular kinematics assessment system with inter and intraobserver reproducibility and normative data for healthy adults.

Ruiz Ibán Miguel Angel MA   Paniagua Gonzalez Andrea A   Muraccini Marco M   Asenjo Gismero Cristina C   Varini Alessandro A   Berardi Antonella A   Mantovani Matteo M  

Journal of experimental orthopaedics 20200513 1


<h4>Purpose</h4>To evaluate the intra and interobserver reproducibility of a new system that assesses the three-dimensional humero-scapulo-thoracic kinematics using wearable technology in an outpatient setting. To obtain normative data with the system for scapular angular motions in three planes.<h4>Methods</h4>The SHoW Motion 3D kinematic tracking system is a motion analysis system that uses wireless wearable non-invasive inertial-magnetic sensors to assess the three-dimensional kinematics of t  ...[more]

Similar Datasets

| S-EPMC8822462 | biostudies-literature
| S-EPMC8714653 | biostudies-literature
| S-EPMC6105013 | biostudies-literature
| S-EPMC2694801 | biostudies-literature
| S-EPMC5527015 | biostudies-literature
| S-EPMC7484087 | biostudies-literature
| S-EPMC10285390 | biostudies-literature
| S-EPMC8584785 | biostudies-literature
| S-EPMC6127158 | biostudies-literature
| S-EPMC5989764 | biostudies-literature