A brief airway occlusion is sufficient to measure the patient's inspiratory effort/electrical activity of the diaphragm index (PEI).


ABSTRACT: Pressure generated by patient's inspiratory muscles (Pmus) during assisted mechanical ventilation is of significant relevance. However, Pmus is not commonly measured since an esophageal balloon catheter is required. We have previously shown that Pmus can be estimated by measuring the electrical activity of the diaphragm (EAdi) through the Pmus/EAdi index (PEI). We investigated whether PEI could be reliably measured by a brief end-expiratory occlusion maneuver to propose an automated PEI measurement performed by the ventilator. Pmus, EAdi, airway pressure (Paw), and flow waveforms of 12 critically ill patients undergoing assisted mechanical ventilation were recorded. Repeated end-expiratory occlusion maneuvers were performed. PEI was measured at 100 ms (PEI0.1) and 200 ms (PEI0.2) from the start of the occlusion and compared to the PEI measured at the maximum Paw deflection (PEIoccl, reference). PEI0.1 and PEI0.2 tightly correlated with PEIoccl, (p?

SUBMITTER: Coppadoro A 

PROVIDER: S-EPMC7223874 | biostudies-literature | 2020 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

A brief airway occlusion is sufficient to measure the patient's inspiratory effort/electrical activity of the diaphragm index (PEI).

Coppadoro Andrea A   Rona Roberto R   Bellani Giacomo G   Foti Giuseppe G  

Journal of clinical monitoring and computing 20200109 1


Pressure generated by patient's inspiratory muscles (Pmus) during assisted mechanical ventilation is of significant relevance. However, Pmus is not commonly measured since an esophageal balloon catheter is required. We have previously shown that Pmus can be estimated by measuring the electrical activity of the diaphragm (EAdi) through the Pmus/EAdi index (PEI). We investigated whether PEI could be reliably measured by a brief end-expiratory occlusion maneuver to propose an automated PEI measurem  ...[more]

Similar Datasets

| S-EPMC8797006 | biostudies-literature
| S-EPMC7908005 | biostudies-literature
| S-EPMC6789230 | biostudies-literature
| S-EPMC8943795 | biostudies-literature
| S-EPMC5630544 | biostudies-literature
| S-EPMC7737589 | biostudies-literature
| S-EPMC7092676 | biostudies-literature
| S-EPMC8686581 | biostudies-literature
| S-EPMC8672539 | biostudies-literature
| S-EPMC8427661 | biostudies-literature