Diagnostic accuracy of the FebriDx host response point-of-care test in patients hospitalised with suspected COVID-19.


ABSTRACT: INTRODUCTION:Management of the COVID-19 pandemic is hampered by long delays associated with centralised laboratory PCR testing. In hospitals this leads to poor patient flow and nosocomial transmission and so rapid, accurate diagnostic tests are urgently required. The FebriDx is a point-of-care test that detects an antiviral host response protein in finger prick blood within 10 min, but its accuracy for the identification of COVID-19 is unknown. METHODS:We performed a real-world diagnostic accuracy study of FebriDx in hospitalised patients during the first wave of the pandemic. Measures of diagnostic accuracy were calculated based on FebriDx results compared to the reference standard of SARS-CoV-2 PCR on combined nose and throat swabs. A multivariable predictive model including FebriDx, age, sex, and clinical characteristics was developed and underwent internal validation. RESULTS:FebriDx was performed on 251 patients and gave a valid result in 248. 118 of 248 (48%) were PCR positive for COVID-19. FebriDx results were available after 10 min compared with 1.7 (1.6 to 2.1) hours with point-of-care PCR testing and 23.4 (17.2 to 31.1) hours with laboratory PCR testing. Sensitivity of FebriDx for the identification of COVID-19 was 93% (110/118; 95% CI 87 to 97%) and specificity was 86% (112/130; 95%CI 79 to 92%). Positive and negative likelihood ratios were 6.73 (95%CI 4.37 to 10.37) and 0.08 (95%CI 0.04 to 0.15) respectively. In the multivariate model age, sex and other clinical features did not contribute significantly to the effect of the FebriDx result in distinguishing patients with and without COVID-19. CONCLUSIONS:During the first wave of the pandemic, FebriDx had high accuracy for the identification of COVID-19 in hospitalised adults and could be deployed as a front door triage tool. TRIAL REGISTRATION:ISRCTN14966673.

SUBMITTER: Clark TW 

PROVIDER: S-EPMC7306108 | biostudies-literature | 2020 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

Diagnostic accuracy of the FebriDx host response point-of-care test in patients hospitalised with suspected COVID-19.

Clark Tristan W TW   Brendish Nathan J NJ   Poole Stephen S   Naidu Vasanth V VV   Mansbridge Christopher C   Norton Nicholas N   Wheeler Helen H   Presland Laura L   Ewings Sean S  

The Journal of infection 20200621 4


<h4>Introduction</h4>Management of the COVID-19 pandemic is hampered by long delays associated with centralised laboratory PCR testing. In hospitals this leads to poor patient flow and nosocomial transmission and so rapid, accurate diagnostic tests are urgently required. The FebriDx is a point-of-care test that detects an antiviral host response protein in finger prick blood within 10 min, but its accuracy for the identification of COVID-19 is unknown.<h4>Methods</h4>We performed a real-world di  ...[more]

Similar Datasets

| S-EPMC8263856 | biostudies-literature
| S-EPMC8014604 | biostudies-literature
| S-EPMC7404380 | biostudies-literature
| S-EPMC7576699 | biostudies-literature
| S-EPMC6656794 | biostudies-literature
| S-EPMC8397079 | biostudies-literature
| S-EPMC8426967 | biostudies-literature
2021-12-09 | GSE189610 | GEO
| S-EPMC8362205 | biostudies-literature
| S-EPMC7498257 | biostudies-literature