Validation of magneto-inertial measuring units for measuring hip joint angles.


ABSTRACT: Camera-based motion capture systems are the current gold standard for motion analysis. However, the use of wireless inertial sensor-based systems is increasing in popularity, largely due to convenient portability. The purpose of this study was to validate the use of wireless inertial sensors for measuring hip joint motion with a functional calibration requiring only one motion (walking) and neutral standing. Data were concurrently collected using a 10-camera motion capture system and a wireless inertial sensor-based system. Hip joint angles were measured for 10 participants during walking, jumping jack, and bilateral squat tasks and for a subset (n?=?5) a jump turn task. Camera-based system hip joint angles were calculated from retro-reflective marker positions and sensor-based system angles were calculated in MATLAB using the sensor output quaternions. Most hip joint angles measured with the sensor-based system were within 6° of angles measured with the camera motion capture system. Accurate measurement of motion outside of a laboratory setting has broad implications for diagnosing movement abnormalities, monitoring sports performance, and assessing rehabilitation progress.

SUBMITTER: Horenstein RE 

PROVIDER: S-EPMC7370255 | biostudies-literature | 2019 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Validation of magneto-inertial measuring units for measuring hip joint angles.

Horenstein Rachel E RE   Lewis Cara L CL   Yan Sherry S   Halverstadt Anne A   Shefelbine Sandra J SJ  

Journal of biomechanics 20190523


Camera-based motion capture systems are the current gold standard for motion analysis. However, the use of wireless inertial sensor-based systems is increasing in popularity, largely due to convenient portability. The purpose of this study was to validate the use of wireless inertial sensors for measuring hip joint motion with a functional calibration requiring only one motion (walking) and neutral standing. Data were concurrently collected using a 10-camera motion capture system and a wireless  ...[more]

Similar Datasets

| S-EPMC7506643 | biostudies-literature
| S-EPMC5507609 | biostudies-literature
| S-EPMC5528837 | biostudies-literature
| S-EPMC9348963 | biostudies-literature
| S-EPMC4600010 | biostudies-literature
| S-EPMC10762015 | biostudies-literature
| S-EPMC7147162 | biostudies-literature
| S-EPMC9204569 | biostudies-literature
| S-EPMC10547676 | biostudies-literature
| S-EPMC7841375 | biostudies-literature