Comparison between the Airgo™ Device and a Metabolic Cart during Rest and Exercise.


ABSTRACT: The aim of this study is to compare the accuracy of Airgo™, a non-invasive wearable device that records breath, with respect to a gold standard. In 21 healthy subjects (10 males, 11 females), four parameters were recorded for four min at rest and in different positions simultaneously by Airgo™ and SensorMedics 2900 metabolic cart. Then, a cardio-pulmonary exercise test was performed using the Erg 800S cycle ergometer in order to test Airgo™'s accuracy during physical effort. The results reveal that the relative error median percentage of respiratory rate was of 0% for all positions at rest and for different exercise intensities, with interquartile ranges between 3.5 (standing position) and 22.4 (low-intensity exercise) breaths per minute. During exercise, normalized amplitude and ventilation relative error medians highlighted the presence of an error proportional to the volume to be estimated. For increasing intensity levels of exercise, Airgo™'s estimate tended to underestimate the values of the gold standard instrument. In conclusion, the Airgo™ device provides good accuracy and precision in the estimate of respiratory rate (especially at rest), an acceptable estimate of tidal volume and minute ventilation at rest and an underestimation for increasing volumes.

SUBMITTER: Antonelli A 

PROVIDER: S-EPMC7412454 | biostudies-literature | 2020 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison between the Airgo™ Device and a Metabolic Cart during Rest and Exercise.

Antonelli Andrea A   Guilizzoni Dario D   Angelucci Alessandra A   Melloni Giulio G   Mazza Federico F   Stanzi Alessia A   Venturino Massimiliano M   Kuller David D   Aliverti Andrea A  

Sensors (Basel, Switzerland) 20200715 14


The aim of this study is to compare the accuracy of Airgo™, a non-invasive wearable device that records breath, with respect to a gold standard. In 21 healthy subjects (10 males, 11 females), four parameters were recorded for four min at rest and in different positions simultaneously by Airgo™ and SensorMedics 2900 metabolic cart. Then, a cardio-pulmonary exercise test was performed using the Erg 800S cycle ergometer in order to test Airgo™'s accuracy during physical effort. The results reveal t  ...[more]

Similar Datasets

| S-EPMC7387751 | biostudies-literature
| S-EPMC6647016 | biostudies-literature
| S-BSST691 | biostudies-other
| S-EPMC5553796 | biostudies-other
| S-EPMC4521405 | biostudies-literature
| S-EPMC1153137 | biostudies-other
| S-EPMC6095739 | biostudies-literature
| S-EPMC5396036 | biostudies-literature
| S-EPMC6506471 | biostudies-literature
| S-EPMC5793265 | biostudies-literature