ABSTRACT: Background and aim
Minimizing endoscopist exposure to bodily fluids is important for reducing the risk of infection transmission. This study investigated the patient-endoscopist vertical distance necessary to minimize an endoscopist's facial exposure to a patient's visible droplets during upper gastrointestinal endoscopy and the ability of a new device to prevent droplets from reaching the endoscopist's face.Methods
A model was developed to simulate a patient experiencing a forceful cough during an upper gastrointestinal endoscopy with a model endoscopist. Fluorescent dye was expelled from the model patient's mouth towards the model endoscopist during simulated coughs; dye adhesion to the model endoscopist's face was evaluated using ultraviolet light. The simulation was repeated with the model patient positioned 70-100 cm above the floor, with and without a barrier to shield the patient's face. The accuracy of the cough simulation model and the relationship between patient-endoscopist vertical distance and endoscopist's facial exposure were evaluated.Results
The flow dynamics of the cough simulation model were similar to that of an actual human cough. There was a significant inverse correlation between the patient-endoscopist vertical distance and the model endoscopist's facial exposure, with positive exposures decreasing from 87% at 70 cm to 0% at 100 cm (P < 0.001). The barrier device prevented facial exposure to droplets at all distances.Conclusions
We found that positioning the patient at least 100 cm below the top of the endoscopist's head or using a barrier device minimized the endoscopist's facial exposure to visible droplets during upper gastrointestinal endoscopy.
SUBMITTER: Suzuki S
PROVIDER: S-EPMC7461247 | biostudies-literature | 2021 Apr
REPOSITORIES: biostudies-literature
Journal of gastroenterology and hepatology 20200825 4
<h4>Background and aim</h4>Minimizing endoscopist exposure to bodily fluids is important for reducing the risk of infection transmission. This study investigated the patient-endoscopist vertical distance necessary to minimize an endoscopist's facial exposure to a patient's visible droplets during upper gastrointestinal endoscopy and the ability of a new device to prevent droplets from reaching the endoscopist's face.<h4>Methods</h4>A model was developed to simulate a patient experiencing a force ...[more]