Unknown

Dataset Information

0

Artifact Rates for 2D Retinal Nerve Fiber Layer Thickness Versus 3D Neuroretinal Rim Thickness Using Spectral-Domain Optical Coherence Tomography.


ABSTRACT: Purpose:To compare the rates of clinically significant artifacts for two-dimensional peripapillary retinal nerve fiber layer (RNFL) thickness versus three-dimensional (3D) neuroretinal rim thickness using spectral-domain optical coherence tomography (SD-OCT). Methods:Only one eye per patient was used for analysis of 120 glaucoma patients and 114 normal patients. For RNFL scans and optic nerve scans, 15 artifact types were calculated per B-scan and per eye. Neuroretinal rim tissue was quantified by the minimum distance band (MDB). Global MDB neuroretinal rim thicknesses were calculated before and after manual deletion of B-scans with artifacts and subsequent automated interpolation. A clinically significant artifact was defined as one requiring manual correction or repeat scanning. Results:Among glaucomatous eyes, artifact rates per B-scan were significantly more common in RNFL scans (61.7%, 74 of 120) compared to B-scans in neuroretinal rim volume scans (20.9%, 1423 of 6820) (95% confidence interval [CI], 31.6-50.0; P < 0.0001). For clinically significant artifact rates per eye, optic nerve scans had significantly fewer artifacts (15.8% of glaucomatous eyes, 13.2% of normal eyes) compared to RNFL scans (61.7% of glaucomatous eyes, 25.4% of normal eyes) (glaucoma group: 95% CI, 34.1-57.5, P < 0.0001; normal group: 95% CI, 1.3-23.3, P = 0.03). Conclusions:Compared to the most commonly used RNFL thickness scans, optic nerve volume scans less frequently require manual correction or repeat scanning to obtain accurate measurements. Translational Relevance:This paper illustrates the potential for 3D OCT algorithms to improve in vivo imaging in glaucoma.

SUBMITTER: Park EA 

PROVIDER: S-EPMC7488619 | biostudies-literature | 2020 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Artifact Rates for 2D Retinal Nerve Fiber Layer Thickness Versus 3D Neuroretinal Rim Thickness Using Spectral-Domain Optical Coherence Tomography.

Park Elli A EA   Tsikata Edem E   Lee Jenny Jyoung JJ   Shieh Eric E   Braaf Boy B   Vakoc Benjamin J BJ   Bouma Brett E BE   de Boer Johannes F JF   Chen Teresa C TC  

Translational vision science & technology 20200910 10


<h4>Purpose</h4>To compare the rates of clinically significant artifacts for two-dimensional peripapillary retinal nerve fiber layer (RNFL) thickness versus three-dimensional (3D) neuroretinal rim thickness using spectral-domain optical coherence tomography (SD-OCT).<h4>Methods</h4>Only one eye per patient was used for analysis of 120 glaucoma patients and 114 normal patients. For RNFL scans and optic nerve scans, 15 artifact types were calculated per B-scan and per eye. Neuroretinal rim tissue  ...[more]

Similar Datasets

| S-EPMC6258318 | biostudies-literature
| S-EPMC3806882 | biostudies-literature
| S-EPMC4968911 | biostudies-literature
| S-EPMC3390182 | biostudies-literature
| S-EPMC7214190 | biostudies-literature
| S-EPMC10645813 | biostudies-literature
| S-EPMC6298691 | biostudies-literature
| S-EPMC4627356 | biostudies-literature
| S-EPMC4698808 | biostudies-literature
| S-EPMC9251790 | biostudies-literature