Comparison of waveform-derived corneal stiffness and stress-strain extensometry-derived corneal stiffness using different cross-linking irradiances: an experimental study with air-puff applanation of ex vivo porcine eyes.


ABSTRACT: PURPOSE:To assess corneal stiffening of standard (S-CXL) and accelerated (A-CXL) cross-linking protocols by dynamic corneal response parameters and corneal bending stiffness (Kc[mean/linear]) derived from Corvis (CVS) Scheimpflug-based tonometry. These investigations were validated by corneal tensile stiffness (K[ts]), derived from stress-strain extensometry in ex vivo porcine eyes. METHODS:Seventy-two fresh-enucleated and de-epithelized porcine eyes were soaked in 0.1% riboflavin solution including 10% dextran for 10 min. The eyes were separated into four groups: controls (n?=?18), S-CXL (intensity in mW/cm2*time in min; 3*30) (n?=?18), A-CXL (9*10) (n?=?18), and A-CXL (18*5) (n?=?18), respectively. CXL was performed using CCL Vario. CVS measurements were performed on all eyes. Subsequently, corneal strips were extracted by a double-bladed scalpel and used for stress-strain measurements. K[ts] was calculated from a force-displacement curve. Mean corneal stiffness (Kc[mean]) and constant corneal stiffness (Kc[linear]) were calculated from raw CVS data. RESULTS:In CVS, biomechanical effects of cross-linking were shown to have a significantly decreased deflection amplitude as well as integrated radius, an increased IOP, and SP A1 (P?

SUBMITTER: Herber R 

PROVIDER: S-EPMC7550374 | biostudies-literature | 2020 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of waveform-derived corneal stiffness and stress-strain extensometry-derived corneal stiffness using different cross-linking irradiances: an experimental study with air-puff applanation of ex vivo porcine eyes.

Herber Robert R   Francis Mathew M   Spoerl Eberhard E   Pillunat Lutz E LE   Raiskup Frederik F   Sinha Roy Abhijit A  

Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie 20200617 10


<h4>Purpose</h4>To assess corneal stiffening of standard (S-CXL) and accelerated (A-CXL) cross-linking protocols by dynamic corneal response parameters and corneal bending stiffness (Kc[mean/linear]) derived from Corvis (CVS) Scheimpflug-based tonometry. These investigations were validated by corneal tensile stiffness (K[ts]), derived from stress-strain extensometry in ex vivo porcine eyes.<h4>Methods</h4>Seventy-two fresh-enucleated and de-epithelized porcine eyes were soaked in 0.1% riboflavin  ...[more]

Similar Datasets

| S-EPMC7726963 | biostudies-literature
| S-EPMC4220297 | biostudies-literature
| S-EPMC8542601 | biostudies-literature
| S-EPMC7065616 | biostudies-literature
| S-EPMC5316890 | biostudies-literature
| S-EPMC7812260 | biostudies-literature
| S-EPMC5085055 | biostudies-literature
| S-EPMC7641398 | biostudies-literature
| S-EPMC7997020 | biostudies-literature
| S-EPMC11019595 | biostudies-literature