ABSTRACT: Purpose
To evaluate the intra-unit (RELINTRA) and inter-unit reliability (RELINTER) of two structurally identical units of the metabolic analyser K5 (COSMED, Rome, Italy) that allows to utilize either breath-by-breath (BBB) or dynamic mixing chamber (DMC) technology.Methods
Identical flow- and gas-signals were transmitted to both K5s that always operated simultaneously either in BBB- or DMC-mode. To assess RELINTRA and RELINTER, a metabolic simulator was applied to simulate four graded levels of respiration. RELINTRA and RELINTER were expressed as typical error (TE%) and Intraclass Correlation Coefficient (ICC). To assess also inter-unit differences via natural respiratory signals, 12 male athletes performed one incremental bike step test each in BBB- and DMC-mode. Inter-un
SUBMITTER: Winkert K