Inter- and intra-unit reliability of the COSMED K5: Implications for multicentric and longitudinal testing.


ABSTRACT: PURPOSE:To evaluate the intra-unit (RELINTRA) and inter-unit reliability (RELINTER) of two structurally identical units of the metabolic analyser K5 (COSMED, Rome, Italy) that allows to utilize either breath-by-breath (BBB) or dynamic mixing chamber (DMC) technology. METHODS:Identical flow- and gas-signals were transmitted to both K5s that always operated simultaneously either in BBB- or DMC-mode. To assess RELINTRA and RELINTER, a metabolic simulator was applied to simulate four graded levels of respiration. RELINTRA and RELINTER were expressed as typical error (TE%) and Intraclass Correlation Coefficient (ICC). To assess also inter-unit differences via natural respiratory signals, 12 male athletes performed one incremental bike step test each in BBB- and DMC-mode. Inter-unit differences within biological testing were expressed as percentages. RESULTS:In BBB, TE% of RELINTRA ranged 0.30-0.67 vs. RELINTER 0.16-1.39 and ICC ranged 0.57-1.00 vs. 0.09-1.00. In DMC, TE% of RELINTRA ranged 0.38-0.90 vs. RELINTER 0.03-0.86 and ICC ranged 0.22-1.00 vs. 0.52-1.00. Mean inter-unit differences ranged -2.30-2.20% (Cohen's ds (ds) 0.13-1.52) for BBB- and -0.55-0.61% (ds 0.00-0.65) for DMC-mode, respectively. Inter-unit differences for [Formula: see text] and RER were significant (p < 0.05) at each step. CONCLUSION:Two structurally identical K5-units demonstrated accurate RELINTRA with TE < 2.0% and similar RELINTER during metabolic simulation. During biological testing, inter-unit differences for [Formula: see text] and RER in BBB-mode were higher than 2% with partially large ES in BBB. Hence, the K5 should be allocated personally wherever possible. Otherwise, e.g. in multicenter studies, a decrease in total reliability needs to be considered especially when the BBB-mode is applied.

SUBMITTER: Winkert K 

PROVIDER: S-EPMC7584194 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature

altmetric image

Publications

Inter- and intra-unit reliability of the COSMED K5: Implications for multicentric and longitudinal testing.

Winkert Kay K   Kamnig Rupert R   Kirsten Johannes J   Steinacker Jürgen M JM   Treff Gunnar G  

PloS one 20201023 10


<h4>Purpose</h4>To evaluate the intra-unit (RELINTRA) and inter-unit reliability (RELINTER) of two structurally identical units of the metabolic analyser K5 (COSMED, Rome, Italy) that allows to utilize either breath-by-breath (BBB) or dynamic mixing chamber (DMC) technology.<h4>Methods</h4>Identical flow- and gas-signals were transmitted to both K5s that always operated simultaneously either in BBB- or DMC-mode. To assess RELINTRA and RELINTER, a metabolic simulator was applied to simulate four  ...[more]

Similar Datasets

| S-EPMC7318309 | biostudies-literature
| S-EPMC6142623 | biostudies-literature
| S-EPMC4426257 | biostudies-literature
| S-EPMC4560445 | biostudies-literature
| S-EPMC4574708 | biostudies-literature
| S-EPMC3552977 | biostudies-literature
| S-EPMC10957586 | biostudies-literature
| S-EPMC6927623 | biostudies-literature
| S-EPMC9177712 | biostudies-literature
| S-EPMC7020600 | biostudies-literature