Accuracy of the Dexcom G6 Glucose Sensor during Aerobic, Resistance, and Interval Exercise in Adults with Type 1 Diabetes.


ABSTRACT: The accuracy of continuous glucose monitoring (CGM) sensors may be significantly impacted by exercise. We evaluated the impact of three different types of exercise on the accuracy of the Dexcom G6 sensor. Twenty-four adults with type 1 diabetes on multiple daily injections wore a G6 sensor. Participants were randomized to aerobic, resistance, or high intensity interval training (HIIT) exercise. Each participant completed two in-clinic 30-min exercise sessions. The sensors were applied on average 5.3 days prior to the in-clinic visits (range 0.6-9.9). Capillary blood glucose (CBG) measurements with a Contour Next meter were performed before and after exercise as well as every 10 min during exercise. No CGM calibrations were performed. The median absolute relative difference (MARD) and median relative difference (MRD) of the CGM as compared with the reference CBG did not differ significantly from the start of exercise to the end exercise across all exercise types (ranges for aerobic MARD: 8.9 to 13.9% and MRD: -6.4 to 0.5%, resistance MARD: 7.7 to 14.5% and MRD: -8.3 to -2.9%, HIIT MARD: 12.1 to 16.8% and MRD: -14.3 to -9.1%). The accuracy of the no-calibration Dexcom G6 CGM was not significantly impacted by aerobic, resistance, or HIIT exercise.

SUBMITTER: Guillot FH 

PROVIDER: S-EPMC7600074 | biostudies-literature | 2020 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Accuracy of the Dexcom G6 Glucose Sensor during Aerobic, Resistance, and Interval Exercise in Adults with Type 1 Diabetes.

Guillot Florian H FH   Jacobs Peter G PG   Wilson Leah M LM   Youssef Joseph El JE   Gabo Virginia B VB   Branigan Deborah L DL   Tyler Nichole S NS   Ramsey Katrina K   Riddell Michael C MC   Castle Jessica R JR  

Biosensors 20200929 10


The accuracy of continuous glucose monitoring (CGM) sensors may be significantly impacted by exercise. We evaluated the impact of three different types of exercise on the accuracy of the Dexcom G6 sensor. Twenty-four adults with type 1 diabetes on multiple daily injections wore a G6 sensor. Participants were randomized to aerobic, resistance, or high intensity interval training (HIIT) exercise. Each participant completed two in-clinic 30-min exercise sessions. The sensors were applied on average  ...[more]

Similar Datasets

| S-EPMC8323182 | biostudies-literature
| S-EPMC7757524 | biostudies-literature
| S-EPMC9310589 | biostudies-literature
| S-EPMC4201312 | biostudies-literature
| S-EPMC4997402 | biostudies-literature
| S-EPMC4667335 | biostudies-literature
| S-EPMC6955453 | biostudies-literature
| S-EPMC6551983 | biostudies-literature
| S-EPMC3594238 | biostudies-literature
| S-EPMC3697744 | biostudies-literature