Unknown

Dataset Information

0

Accuracy of UK Rapid Test Consortium (UK-RTC) "AbC-19 Rapid Test" for detection of previous SARS-CoV-2 infection in key workers: test accuracy study.


ABSTRACT: OBJECTIVE:To assess the accuracy of the AbC-19 Rapid Test lateral flow immunoassay for the detection of previous severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infection. DESIGN:Test accuracy study. SETTING:Laboratory based evaluation. PARTICIPANTS:2847 key workers (healthcare staff, fire and rescue officers, and police officers) in England in June 2020 (268 with a previous polymerase chain reaction (PCR) positive result (median 63 days previously), 2579 with unknown previous infection status); and 1995 pre-pandemic blood donors. MAIN OUTCOME MEASURES:AbC-19 sensitivity and specificity, estimated using known negative (pre-pandemic) and known positive (PCR confirmed) samples as reference standards and secondly using the Roche Elecsys anti-nucleoprotein assay, a highly sensitive laboratory immunoassay, as a reference standard in samples from key workers. RESULTS:Test result bands were often weak, with positive/negative discordance by three trained laboratory staff for 3.9% of devices. Using consensus readings, for known positive and negative samples sensitivity was 92.5% (95% confidence interval 88.8% to 95.1%) and specificity was 97.9% (97.2% to 98.4%). Using an immunoassay reference standard, sensitivity was 94.2% (90.7% to 96.5%) among PCR confirmed cases but 84.7% (80.6% to 88.1%) among other people with antibodies. This is consistent with AbC-19 being more sensitive when antibody concentrations are higher, as people with PCR confirmation tended to have more severe disease whereas only 62% (218/354) of seropositive participants had had symptoms. If 1 million key workers were tested with AbC-19 and 10% had actually been previously infected, 84?700 true positive and 18?900 false positive results would be projected. The probability that a positive result was correct would be 81.7% (76.8% to 85.8%). CONCLUSIONS:AbC-19 sensitivity was lower among unselected populations than among PCR confirmed cases of SARS-CoV-2, highlighting the scope for overestimation of assay performance in studies involving only PCR confirmed cases, owing to "spectrum bias." Assuming that 10% of the tested population have had SARS-CoV-2 infection, around one in five key workers testing positive with AbC-19 would be false positives. STUDY REGISTRATION:ISRCTN 56609224.

SUBMITTER: Mulchandani R 

PROVIDER: S-EPMC7656121 | biostudies-literature | 2020 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Accuracy of UK Rapid Test Consortium (UK-RTC) "AbC-19 Rapid Test" for detection of previous SARS-CoV-2 infection in key workers: test accuracy study.

Mulchandani Ranya R   Jones Hayley E HE   Taylor-Phillips Sian S   Shute Justin J   Perry Keith K   Jamarani Shabnam S   Brooks Tim T   Charlett Andre A   Hickman Matthew M   Oliver Isabel I   Kaptoge Stephen S   Danesh John J   Di Angelantonio Emanuele E   Ades Anthony E AE   Wyllie David H DH  

BMJ (Clinical research ed.) 20201111


<h4>Objective</h4>To assess the accuracy of the AbC-19 Rapid Test lateral flow immunoassay for the detection of previous severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infection.<h4>Design</h4>Test accuracy study.<h4>Setting</h4>Laboratory based evaluation.<h4>Participants</h4>2847 key workers (healthcare staff, fire and rescue officers, and police officers) in England in June 2020 (268 with a previous polymerase chain reaction (PCR) positive result (median 63 days previously), 257  ...[more]

Similar Datasets

| S-EPMC8782985 | biostudies-literature
| S-EPMC7997203 | biostudies-literature
| S-EPMC7534827 | biostudies-literature
| S-EPMC8118553 | biostudies-literature
| S-EPMC8263628 | biostudies-literature
| S-EPMC8163726 | biostudies-literature
| S-EPMC8513858 | biostudies-literature
| S-EPMC6393825 | biostudies-literature
| S-EPMC7404380 | biostudies-literature
| S-EPMC7881288 | biostudies-literature