Doppler ultrasound cardiac gating of intracranial flow at 7T.


ABSTRACT:

Background

Ultra-high field magnetic resonance imaging (MR) may be used to improve intracranial blood flow measurements. However, standard cardiac synchronization methods tend to fail at ultra-high field MR. Therefore, this study aims to investigate an alternative synchronization technique using Doppler ultrasound.

Methods

Healthy subjects (n?=?9) were examined with 7T MR. Flow was measured in the M1-branch of the middle cerebral artery (MCA) and in the cerebral aqueduct (CA) using through-plane phase contrast (2D flow). Flow in the circle of Willis was measured with three-dimensional, three-directional phase contrast (4D flow). Scans were gated with Doppler ultrasound (DUS) and electrocardiogram (ECG), and pulse oximetry data (POX) was collected simultaneously. False negative and false positive trigger events were counted for ECG, DUS and POX, and quantitative flow measures were compared.

Results

There were fewer false positive triggers for DUS compared to ECG (5.3?±?11 vs. 25?±?31, p?=?0.031), while no other measured parameters differed significantly. Net blood flow in M1 was similar between DUS and ECG for 2D flow (1.5?±?0.39 vs. 1.6?±?0.41, bias?±?1.96SD: -?0.021?±?0.36) and 4D flow (1.8?±?0.48 vs. 9?±?0.59, bias?±?1.96SD: -?0.086?±?0.57 ml). Net CSF flow per heart beat in the CA was also similar for DUS and ECG (3.6?±?2.1 vs. 3.0?±?5.8, bias?±?1.96SD: 0.61?±?13.6 ?l).

Conclusion

Gating with DUS produced fewer false trigger events than using ECG, with similar quantitative flow values. DUS gating is a promising technique for cardiac synchronization at 7T.

SUBMITTER: Markenroth Bloch K 

PROVIDER: S-EPMC7724705 | biostudies-literature | 2020 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Doppler ultrasound cardiac gating of intracranial flow at 7T.

Markenroth Bloch Karin K   Kording Fabian F   Töger Johannes J  

BMC medical imaging 20201209 1


<h4>Background</h4>Ultra-high field magnetic resonance imaging (MR) may be used to improve intracranial blood flow measurements. However, standard cardiac synchronization methods tend to fail at ultra-high field MR. Therefore, this study aims to investigate an alternative synchronization technique using Doppler ultrasound.<h4>Methods</h4>Healthy subjects (n = 9) were examined with 7T MR. Flow was measured in the M1-branch of the middle cerebral artery (MCA) and in the cerebral aqueduct (CA) usin  ...[more]

Similar Datasets

| S-EPMC9363520 | biostudies-literature
| S-EPMC8774649 | biostudies-literature
| S-EPMC9935734 | biostudies-literature
| S-EPMC8019737 | biostudies-literature
| S-EPMC5947608 | biostudies-literature
| S-EPMC8550238 | biostudies-literature
| S-EPMC6488013 | biostudies-literature
| S-EPMC10250020 | biostudies-literature
| S-EPMC9969216 | biostudies-literature
| S-EPMC9495671 | biostudies-literature