Quantitative ultrasound radiofrequency data analysis for the assessment of hepatic steatosis using the controlled attenuation parameter as a reference standard.


ABSTRACT:

Purpose

This study was aimed to investigate the value of quantitative ultrasound (US) parameters from radiofrequency (RF) data analysis for assessing hepatic steatosis, using controlled attenuation parameter (CAP)-based steatosis grades as the reference standard.

Methods

We analyzed 243 participants with both B-mode liver US with RF data acquisition and CAP measurements. On B-mode US images, hepatic steatosis was visually scored (0/1/2/3, none/mild/moderate/severe), and the hepatorenal index (HRI) was calculated. From the RF data analysis, the tissue scatter-distribution imaging parameter (TSI-p) and tissue attenuation imaging parameter (TAI-p) of the liver parenchyma were measured. US parameters were correlated with CAP-based steatosis grades (S0/1/2/3, none/mild/moderate/severe) and their diagnostic performance was evaluated using receiver operating characteristic (ROC) curve analysis. Multivariate linear regression analysis was performed to identify determinants of TSI-p and TAI-p.

Results

Participants were classified as having S0 (n=152), S1 (n=54), S2 (n=14), and S3 (n=23) on CAP measurements. TSI-p and TAI-p were significantly correlated with steatosis grades (? =0.593 and ?=-0.617, P<0.001 for both). For predicting ?S1, ?S2, and S3, the areas under the ROC curves (AUCs) of TSI-p were 0.827/0.914/0.917; TAI-p, 0.844/0.914/0.909; visual scores, 0.659/0.778/0.794; and HRI, 0.629/0.751/0.759, respectively. TSI-p and TAI-p had significantly higher AUCs than did visual scores or HRI for ?S1 or ?S2 (P?0.003). In the multivariate analysis, the transient elastography-based fibrosis grade (P=0.034) and steatosis grade (P<0.001) were independent determinants of TSI-p, while steatosis grade (P<0.001) was an independent determinant of TAI-p.

Conclusion

TSI-p and TAI-p derived from US RF data may be useful for detecting hepatic steatosis and assessing its severity.

SUBMITTER: Jeon SK 

PROVIDER: S-EPMC7758108 | biostudies-literature | 2021 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Quantitative ultrasound radiofrequency data analysis for the assessment of hepatic steatosis using the controlled attenuation parameter as a reference standard.

Jeon Sun Kyung SK   Joo Ijin I   Kim So Yeon SY   Jang Jong Keon JK   Park Juil J   Park Hee Sun HS   Lee Eun Sun ES   Lee Jeong Min JM  

Ultrasonography (Seoul, Korea) 20200509 1


<h4>Purpose</h4>This study was aimed to investigate the value of quantitative ultrasound (US) parameters from radiofrequency (RF) data analysis for assessing hepatic steatosis, using controlled attenuation parameter (CAP)-based steatosis grades as the reference standard.<h4>Methods</h4>We analyzed 243 participants with both B-mode liver US with RF data acquisition and CAP measurements. On B-mode US images, hepatic steatosis was visually scored (0/1/2/3, none/mild/moderate/severe), and the hepato  ...[more]

Similar Datasets

| S-EPMC8038574 | biostudies-literature
| S-EPMC8236371 | biostudies-literature
| S-EPMC8118256 | biostudies-literature
| S-EPMC9387959 | biostudies-literature
| S-EPMC10071056 | biostudies-literature
| S-EPMC9011258 | biostudies-literature
| S-EPMC5867216 | biostudies-literature
| S-EPMC3956815 | biostudies-literature
| S-EPMC5659780 | biostudies-literature
| S-EPMC10697075 | biostudies-literature