Unknown

Dataset Information

0

Comparison of phase-resolved functional lung (PREFUL) MRI derived perfusion and ventilation parameters at 1.5T and 3T in healthy volunteers.


ABSTRACT:

Purpose

The purpose of this study is to evaluate the influence of different field strengths on perfusion and ventilation parameters, SNR and CNR derived by PREFUL MRI using predefined sequence parameters.

Methods

Data sets of free breathing 2d FLASH lung MRI were acquired from 15 healthy subjects at 1.5T and 3T (Magnetom Avanto and Skyra, Siemens Healthcare, Erlangen, Germany) with a maximum period of 3 days in between. The processed functional parameters regional ventilation (RVent), perfusion (Q), quantified perfusion (QQuant), perfusion defect percentage (QDP), ventilation defect percentage (VDP) and ventilation-perfusion match (VQM) were compared for systematic differences. Signal- and contrast-to-noise ratio (SNR and CNR) of both acquisitions were analyzed.

Results

RVent, Q, VDP, SNR and CNR presented no significant differences between 1.5T and 3T. QQuant (1.5T vs. 3T, P = 0.04), and QDP (1.5T vs. 3T, P?0.01) decreased significantly at 3T. Consequently, VQM increased significantly (1.5T vs. 3T, P?0.01). Skewness and kurtosis of the Q-values increased significantly at 3T (P?0.01). The mean Sørensen-Dice coefficients between both series were 0.91 for QDP and 0.94 for VDP. The Bland-Altman analysis of both series showed mean differences of 4.29% for QDP, 1.23% for VDP and -5.15% for VQM. Using the above-mentioned parameters for three-day repeatability at two different scanners and field strengths, the retrospective power calculation showed, that a sample size of 15 can detect differences of 3.7% for QDP, of 2.9% for VDP and differences of 2.6% for VQM.

Conclusion

Significant differences in QDP may be related to field inhomogeneities, which is expressed by increasing skewness and kurtosis at 3T. QQuant reveals only poor reproducibility between 1.5T and 3T. RVent, Q, VDP, SNR and CNR were not altered significantly at the used sequence parameters. Healthy participants with minimal defects present high spatial agreement of QDP and VDP.

SUBMITTER: Glandorf J 

PROVIDER: S-EPMC7773267 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of phase-resolved functional lung (PREFUL) MRI derived perfusion and ventilation parameters at 1.5T and 3T in healthy volunteers.

Glandorf Julian J   Klimeš Filip F   Voskrebenzev Andreas A   Gutberlet Marcel M   Behrendt Lea L   Crisosto Cristian C   Wacker Frank F   Ciet Pierluigi P   Wild Jim M JM   Vogel-Claussen Jens J  

PloS one 20201230 12


<h4>Purpose</h4>The purpose of this study is to evaluate the influence of different field strengths on perfusion and ventilation parameters, SNR and CNR derived by PREFUL MRI using predefined sequence parameters.<h4>Methods</h4>Data sets of free breathing 2d FLASH lung MRI were acquired from 15 healthy subjects at 1.5T and 3T (Magnetom Avanto and Skyra, Siemens Healthcare, Erlangen, Germany) with a maximum period of 3 days in between. The processed functional parameters regional ventilation (RVe  ...[more]

Similar Datasets

| S-EPMC8277031 | biostudies-literature
| S-EPMC4911168 | biostudies-literature
| S-EPMC7021666 | biostudies-literature
| S-EPMC5904463 | biostudies-literature
| S-EPMC8996400 | biostudies-literature
| S-EPMC6004122 | biostudies-literature
| S-EPMC9163566 | biostudies-literature
| S-EPMC3166692 | biostudies-literature
| S-EPMC8795032 | biostudies-literature
| S-EPMC6812822 | biostudies-literature