Unknown

Dataset Information

0

Early clinical experience with a total body irradiation technique using field-in-field beams and on-line image guidance.


ABSTRACT:

Background and purpose

Total body irradiation (TBI) is a treatment used in the conditioning of patients prior to hematopoietic stem cell transplantation. We developed an extended-distance TBI technique using a conventional linac with multi-leaf collimator to deliver a homogeneous dose, and spare critical organs.

Materials and methods

Patients were treated either in lateral recumbent or in supine position depending on the dose level. A conventional linac was used with the patient midline at 350 cm from the beam source. A series of beams was prepared manually using a 3D treatment planning system (TPS) aiming to improve dose homogeneity, spare the organs at risk and facilitate accurate patient positioning. An optimized dose calculation model for extended-distance treatments was developed using phantom measurements. During treatment, in-vivo dosimetry was performed using electronic dosimeters, and accurate positioning was verified using a mobile megavoltage imager. We analyzed dose volume histogram parameters for 19 patients, and in-vivo measurements for 46 delivered treatment fractions.

Results

Optimization of the dose calculation model for TBI improved dose calculation by 2.1% at the beam axis, and 17% at the field edge. Treatment planning dose objectives and constraints were met for 16 of 19 patients. Results of in-vivo dosimetry were within the set limitations (±10%) with mean deviations of 3.7% posterior of the lungs and 0.6% for the abdomen.

Conclusions

We developed a TBI treatment technique using a conventional linac and TPS that can reliably be used in the conditioning regimen of patients prior to stem cell transplantation.

SUBMITTER: van Leeuwen RGH 

PROVIDER: S-EPMC7807619 | biostudies-literature | 2020 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

Early clinical experience with a total body irradiation technique using field-in-field beams and on-line image guidance.

van Leeuwen Ruud G H RGH   Verwegen Drean D   van Kollenburg Peter G M PGM   Swinkels Marc M   van der Maazen Richard W M RWM  

Physics and imaging in radiation oncology 20201001


<h4>Background and purpose</h4>Total body irradiation (TBI) is a treatment used in the conditioning of patients prior to hematopoietic stem cell transplantation. We developed an extended-distance TBI technique using a conventional linac with multi-leaf collimator to deliver a homogeneous dose, and spare critical organs.<h4>Materials and methods</h4>Patients were treated either in lateral recumbent or in supine position depending on the dose level. A conventional linac was used with the patient m  ...[more]

Similar Datasets

| S-EPMC7033804 | biostudies-literature
| S-EPMC7807637 | biostudies-literature
| S-EPMC6937701 | biostudies-literature
2010-06-09 | E-GEOD-1366 | biostudies-arrayexpress
2004-06-17 | GSE1366 | GEO
| S-EPMC6839384 | biostudies-literature
| S-EPMC7674696 | biostudies-literature