Multi-analyser detector (MAD) for high-resolution and high-energy powder X-ray diffraction.


ABSTRACT: For high-resolution powder diffraction in material science, high photon energies are necessary, especially for in situ and in operando experiments. For this purpose, a multi-analyser detector (MAD) was developed for the high-energy beamline P02.1 at PETRA III of the Deutsches Elektronen-Synchrotron (DESY). In order to be able to adjust the detector for the high photon energies of 60?keV, an individually adjustable analyser-crystal setup was designed. The adjustment is performed via piezo stepper motors for each of the ten channels. The detector shows a low and flat background as well as a high signal-to-noise ratio. A range of standard materials were measured for characterizing the performance. Two exemplary experiments were performed to demonstrate the potential for sophisticated structural analysis with the MAD: (i) the structure of a complex material based on strontium niobate titanate and strontium niobate zirconate was determined and (ii) an in situ stroboscopy experiment with an applied electric field on a highly absorbing piezoceramic was performed. These experiments demonstrate the capabilities of the new MAD, which advances the frontiers of the structural characterization of materials.

SUBMITTER: Schokel A 

PROVIDER: S-EPMC7842216 | biostudies-literature | 2021 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Multi-analyser detector (MAD) for high-resolution and high-energy powder X-ray diffraction.

Schökel Alexander A   Etter Martin M   Berghäuser Andreas A   Horst Alexander A   Lindackers Dirk D   Whittle Thomas A TA   Schmid Siegbert S   Acosta Matias M   Knapp Michael M   Ehrenberg Helmut H   Hinterstein Manuel M  

Journal of synchrotron radiation 20210101 Pt 1


For high-resolution powder diffraction in material science, high photon energies are necessary, especially for in situ and in operando experiments. For this purpose, a multi-analyser detector (MAD) was developed for the high-energy beamline P02.1 at PETRA III of the Deutsches Elektronen-Synchrotron (DESY). In order to be able to adjust the detector for the high photon energies of 60 keV, an individually adjustable analyser-crystal setup was designed. The adjustment is performed via piezo stepper  ...[more]

Similar Datasets

| S-EPMC8366427 | biostudies-literature
| S-EPMC4416682 | biostudies-literature
| S-EPMC10481261 | biostudies-literature
| S-EPMC2924792 | biostudies-other
| S-EPMC9901925 | biostudies-literature
| S-EPMC9023493 | biostudies-literature
| S-EPMC5571747 | biostudies-literature
| S-EPMC7312157 | biostudies-literature
| S-EPMC5414395 | biostudies-literature
| S-EPMC6706080 | biostudies-literature