Multicenter evaluation of the Panbio™ COVID-19 rapid antigen-detection test for the diagnosis of SARS-CoV-2 infection.


ABSTRACT: The standard RT-PCR assay for coronavirus disease 2019 (COVID-19) is laborious and time-consuming, limiting testing availability. Rapid antigen-detection tests are faster and less expensive; however, the reliability of these tests must be validated before they can be used widely. The objective of this study was to determine the performance of the Panbio™ COVID-19 Ag Rapid Test Device (PanbioRT) (Abbott) in detecting severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in nasopharyngeal swab specimens. This prospective multicentre study was carried out in ten Spanish university hospitals and included individuals with clinical symptoms or epidemiological criteria of COVID-19. Only individuals with ≤7 days from the onset of symptoms or from exposure to a confirmed case of COVID-19 were included. Two nasopharyngeal samples were taken to perform the PanbioRT as a point-of-care test and a diagnostic RT-PCR test. Among the 958 patients studied, 325 (90.5%) had true-positive results. The overall sensitivity and specificity for the PanbioRT were 90.5% (95%CI 87.5-93.6) and 98.8% (95%CI 98-99.7), respectively. Sensitivity in participants who had a threshold cycle (CT) < 25 for the RT-PCR test was 99.5% (95%CI 98.4-100), and in participants with ≤5 days of the clinical course it was 91.8% (95%CI 88.8-94.8). Agreement between techniques was 95.7% (κ score 0.90; 95%CI 0.88-0.93). The PanbioRT performs well clinically, with even more reliable results for patients with a shorter clinical course of the disease or a higher viral load. The results must be interpreted based on the local epidemiological context.

SUBMITTER: Merino P 

PROVIDER: S-EPMC7884234 | biostudies-literature | 2021 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Multicenter evaluation of the Panbio™ COVID-19 rapid antigen-detection test for the diagnosis of SARS-CoV-2 infection.

Merino Paloma P   Guinea Jesús J   Muñoz-Gallego Irene I   González-Donapetry Patricia P   Galán Juan Carlos JC   Antona Nerea N   Cilla Gustavo G   Hernáez-Crespo Silvia S   Díaz-de Tuesta José Luis JL   Gual-de Torrella Ana A   González-Romo Fernando F   Escribano Pilar P   Sánchez-Castellano Miguel Ángel MÁ   Sota-Busselo Mercedes M   Delgado-Iribarren Alberto A   García Julio J   Cantón Rafael R   Muñoz Patricia P   Folgueira María Dolores MD   Cuenca-Estrella Manuel M   Oteo-Iglesias Jesús J  

Clinical microbiology and infection : the official publication of the European Society of Clinical Microbiology and Infectious Diseases 20210216


<h4>Objectives</h4>The standard RT-PCR assay for coronavirus disease 2019 (COVID-19) is laborious and time-consuming, limiting testing availability. Rapid antigen-detection tests are faster and less expensive; however, the reliability of these tests must be validated before they can be used widely. The objective of this study was to determine the performance of the Panbio™ COVID-19 Ag Rapid Test Device (PanbioRT) (Abbott) in detecting severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2)  ...[more]

Similar Datasets

| S-EPMC7833843 | biostudies-literature
| S-EPMC7881288 | biostudies-literature
| S-EPMC9289906 | biostudies-literature
| S-EPMC8526115 | biostudies-literature
| S-EPMC9514621 | biostudies-literature
| S-EPMC7662075 | biostudies-literature
| S-EPMC9955641 | biostudies-literature
| S-EPMC9876661 | biostudies-literature
| S-EPMC8224876 | biostudies-literature
| S-EPMC8629250 | biostudies-literature