ABSTRACT: Background
Enzyme-linked immunosorbent assay (ELISA) has traditionally been used to detect myeloperoxidase (MPO) and proteinase 3 (PR3) antibodies, although it is time-consuming and physically demanding. As a novel and highly effective immunoassay, we compared chemiluminescent immunoassay (CIA) with ELISA to verify the application value of CIA in MPO and PR3 antibodies detection.Methods
By ELISA and CIA, serum levels of anti-MPO and anti-PR3 antibodies were measured in 63 anti-neutrophil cytoplasmic antibody (ANCA)-associated vasculitis (AAV) patients (AAV group), including 47 microscopic polyangiitis (MPA) patients and 16 granulomatosis with polyangiitis (GPA) patients, in addition, 68 patients in interference control group (IC group), 19 healthy subjects in healthy control group (HC group). We compared MPO and PR3 antibodies levels and positive rates measured by these two methods among groups. Relationship and coincidence rate between ELISA and CIA were investigated. Diagnostic values for clinical outcomes for MPO and PR3 antibodies were assessed by receiver operator characteristic (ROC) curve.Results
In AAV patients, when detecting anti-MPO (r = .90) and anti-PR3 (r = .81), CIA was highly correlated with ELISA, companying with highly total (88.89%, 92.06%, respectively) and positive coincidence rates (84.78%, 77.27%, respectively). In HC group, anti-PR3 positive rate detected by both immunoassay were 0, anti-MPO almost were 0, which without statistically significant difference (P = .32). In IC group, the total (76.47%, 58.82, respectively) and positive coincidence rates (48.38%, 30.00%, respectively) of anti-MPO and anti-PR3 were the lowest, but the negative coincidence rates reached 100%. By CIA, similar to ELISA, the levels of anti-MPO were significantly higher both in AAV patients (56.00; [4.40-235.30]) and MPA patients (98.00; [27.90-324.70]) compared with either IC group (3.20; [3.20-18.55) (P < .0001) or HC group (3.20; [3.20-3.20]) (P < .0001), yielded an area under curve (AUC) of 0.76 for AAV and 0.89 for MPA, the concentration of anti-PR3 in GPA group (66.65; [24.43-150.00]) was significantly higher than that in IC group (2.3; [2.3-10.95]) (P < .0001) and HC group (2.3; [2.3-2.3]) (P < .0001), with an AUC of 0.92.Conclusion
Similar to ELISA, CIA was competent to detect MPO and PR3 antibodies in AAV patients and healthy population, thus distinguish AAV patients from IC group and HC group and effectively diagnose MPA and GPA.
SUBMITTER: Hou X
PROVIDER: S-EPMC7891508 | biostudies-literature | 2021 Feb
REPOSITORIES: biostudies-literature
Journal of clinical laboratory analysis 20201009 2
<h4>Background</h4>Enzyme-linked immunosorbent assay (ELISA) has traditionally been used to detect myeloperoxidase (MPO) and proteinase 3 (PR3) antibodies, although it is time-consuming and physically demanding. As a novel and highly effective immunoassay, we compared chemiluminescent immunoassay (CIA) with ELISA to verify the application value of CIA in MPO and PR3 antibodies detection.<h4>Methods</h4>By ELISA and CIA, serum levels of anti-MPO and anti-PR3 antibodies were measured in 63 anti-ne ...[more]