Unknown

Dataset Information

0

Multi-organ comparison of flow-based arterial spin labeling techniques: Spatially non-selective labeling for cerebral and renal perfusion imaging.


ABSTRACT:

Purpose

Flow-based arterial spin labeling (ASL) techniques provide a transit-time insensitive alternative to the more conventional spatially selective ASL techniques. However, it is not clear which flow-based ASL technique performs best and also, how these techniques perform outside the brain (taking into account eg, flow-dynamics, field-inhomogeneity, and organ motion). In the current study we aimed to compare 4 flow-based ASL techniques (ie, velocity selective ASL, acceleration selective ASL, multiple velocity selective saturation ASL, and velocity selective inversion prepared ASL [VSI-ASL]) to the current spatially selective reference techniques in brain (ie, pseudo-continuous ASL [pCASL]) and kidney (ie, pCASL and flow alternating inversion recovery [FAIR]).

Methods

Brain (n = 5) and kidney (n = 6) scans were performed in healthy subjects at 3T. Perfusion-weighted signal (PWS) maps were generated and ASL techniques were compared based on temporal SNR (tSNR), sensitivity to perfusion changes using a visual stimulus (brain) and robustness to respiratory motion by comparing scans acquired in paced-breathing and free-breathing (kidney).

Results

In brain, all flow-based ASL techniques showed similar tSNR as pCASL, but only VSI-ASL showed similar sensitivity to perfusion changes. In kidney, all flow-based ASL techniques had comparable tSNR, although all lower than FAIR. In addition, VSI-ASL showed a sensitivity to B1 -inhomogeneity. All ASL techniques were relatively robust to respiratory motion.

Conclusion

In both brain and kidney, flow-based ASL techniques provide a planning-free and transit-time insensitive alternative to spatially selective ASL techniques. VSI-ASL shows the most potential overall, showing similar performance as the golden standard pCASL in brain. However, in kidney, a reduction of B1 -sensitivity of VSI-ASL is necessary to match the performance of FAIR.

SUBMITTER: Franklin SL 

PROVIDER: S-EPMC7898485 | biostudies-literature | 2021 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Multi-organ comparison of flow-based arterial spin labeling techniques: Spatially non-selective labeling for cerebral and renal perfusion imaging.

Franklin Suzanne L SL   Bones Isabell K IK   Harteveld Anita A AA   Hirschler Lydiane L   van Stralen Marijn M   Qin Qin Q   de Boer Anneloes A   Hoogduin Johannes M JM   Bos Clemens C   van Osch Matthias J P MJP   Schmid Sophie S  

Magnetic resonance in medicine 20201130 5


<h4>Purpose</h4>Flow-based arterial spin labeling (ASL) techniques provide a transit-time insensitive alternative to the more conventional spatially selective ASL techniques. However, it is not clear which flow-based ASL technique performs best and also, how these techniques perform outside the brain (taking into account eg, flow-dynamics, field-inhomogeneity, and organ motion). In the current study we aimed to compare 4 flow-based ASL techniques (ie, velocity selective ASL, acceleration selecti  ...[more]

Similar Datasets

| S-EPMC4369335 | biostudies-literature
| S-EPMC10841663 | biostudies-literature
| S-EPMC3185150 | biostudies-literature
| S-EPMC7402012 | biostudies-literature
| S-EPMC3087466 | biostudies-literature
| S-EPMC6615405 | biostudies-literature
| S-EPMC6598714 | biostudies-literature
| S-EPMC3769358 | biostudies-literature
| S-EPMC6311674 | biostudies-literature
| S-EPMC6899879 | biostudies-literature